AEC(Automatic Exposure Controls)

¡Supera tus tareas y exámenes ahora con Quizwiz!

density

Two outside detectors activated when __________ of the body part over the 2 outside cells is close to the same density

brightness increase decrease

Optimal kVp is crucial: If kVp is changed, the AEC will compensate with an adjustment in exposure time to adjust __________ If kVp is decreased, the exposure time will automatically __________ If kVp is increased, the exposure time will automatically __________

three anatomy

Photocell/Sensors/Detectors: The radiographer selects the configuration of these devices, determining which one(s) of the ______ actually measures radiation exposure reaching the IR. Configuration is dependent on the ________ of interest

detectors; visible light photomultiplier; photodiodes large

Phototimers: Light paddles, coated with a fluorescent material, serve as the ________, and the radiation interacts with the paddles, producing what? This light is transmitted to remote ___________ tubes or ___________ that convert this light into electricity. The timer is tripped, and the radiographic exposure is terminated when a sufficiently ______ charge has been received.

Entrance or Exit Type kVp; mA Time

Position of detectors varies in 2 ways: Regardless of type, the technologist still controls the ____ and ___ ______ is controlled by AEC device

patient; electrical termination

Radiation is transmitted through the _________ and converted into an __________ signal, terminating the exposure time. A predetermined level of radiation must be reached before exposure ____________.

radiation exposure

Summary: Since visual cues lending to increased or decreased radiographic density with film/screen are not as prominent in DR, the radiographer must be very conscientious about ___________ ___________ to the patient.

Entrance Exit

The _________ type is positioned between the patient and the IR. The most common type. The ________ type is positioned after the patient and image receptor.

Outer

Which detectors do you activate for a Abdomen: Supine/Oblique/ Lateral/Upright?

All

Which detectors do you activate for a BE: AP/PA/RAO/LAO/RPO/LPO/Decub?

Center

Which detectors do you activate for a BE: Lateral?

Center

Which detectors do you activate for a Clavicle: AP/Axial?

Center

Which detectors do you activate for a Femur: AP/Lateral?

Center

Which detectors do you activate for a Hip: AP/Frog Leg Lateral?

Outer

Which detectors do you activate for a IVU?

Center

Which detectors do you activate for a Knee: AP, Oblique, Lateral?

Center

Which detectors do you activate for a Lateral Chest?

Outer

Which detectors do you activate for a PA or AP Chest?

Center

Which detectors do you activate for a Ribs: AP/PA bilateral; RAO/LAO/RPO/LPO; Lower?

Outer

Which detectors do you activate for a Ribs: AP/PA; bilateral?

Center

Which detectors do you activate for a SI Joints: AP/Obliques?

Center

Which detectors do you activate for a Shoulder: AP/Grashey?

All

Which detectors do you activate for a Small Bowel?

Center

Which detectors do you activate for a Sternum: RAO, LPO, R/L Lateral?

Center

Which detectors do you activate for a Stomach: AP/PA/RAO/LPO/R-Lateral?

Center

Which detectors do you activate for a T-Spine, L-Spine, C-Spine?

Center

Which detectors do you activate for a Tib/Fib: AP/Lateral?

Center

Which detectors do you activate for an Esophagus: RAO/AP/Lateral?

Automatic Exposure Controls

A system used to consistently control the amount of radiation reaching the image receptor by terminating the length of exposure. Began as early as 1929

exposure time brightness kVp adjusts noise decrease

AEC devices affect only what? Recall that time controls radiographic ____________, as it is part of mAs ______ influences contrast and affects radiographic brightness If a digital IR is underexposed or overexposed, the computer ________ for the exposure error, but the image quality and/or patient exposure is affected. Underexposure may result in the visibility of quantum _______, and overexposure will increase patient exposure and may __________ contrast.

electricity Phototimer & Ionization Chamber Systems

AEC systems differ by the device used to convert radiation into __________ What are the two AEC systems?

NOT DIAGNOSTIC

Activating the wrong detector will result in a image that is what?

Consistent Reduced efficiency or workload Automatic

Advantages of AEC: ____________ brightness on the images regardless of patient size and thickness. ____________ patient dose, since most exposures are correct and no need for repeats. Increase in departmental _________ or ___________, because of not having as many repeats. _____________ compensation for exposure values; even if there is an SID change.

decreases

Anatomic part has excessive gas Exposure time ___________

time BACK UP TIMER 600 mAs

Backup Timer: Recall: AEC controls the exposure _____- determined by how long it takes the exit radiation to charge the x ray timer to terminate The exposure time set on the control panel is the WHAT? *modern control panels may be automatically selecting the backup timer (at a max of ______ mAs). the following explanation is still crucial regarding AEC function

longer 1.5 600 mAs

Backup Timer: The backup timer must be set ___________ than the length of the expected exposure time (and longer than the min. response time) Typically _____ times more that what is required for the manual exposure factors for the anatomical part. FDA requires that the back up time be set as to never exceed _____ mAs

centered or unilateral

Center detectors activated when anatomy is __________ or __________ in interest

foreign bodies exposure IR collimation QC

Conditions that may cause AEC to malfunction: Do not cover detector with any ________ _________(lead, prosthesis) Selection of wrong bucky --most modern systems with sensors will not allow an ________ if an IR is not in the activated bucky Changes in ____ type --systems are calibrated for specific IRs and should not be changed Errors in _____________ --no collimation: increase scatter: premature termination --over collimation: prolonged exposure required: increase dose Neglect in _____ protocols

Detector IR Collimation Density

Film Screen: -Effect on film density- ____________ combination Part placement not covering detector(s) Type of ____ AEC calibrated to one type of image receptor ______________ Excessive or insufficient may affect exposure time _______ control Predetermined level of radiation adjusted

mA kVp SID OID

Film Screen: -No effect on film density- ______ Exposure time adjusted ______ Exposure time/contrast adjusted ______ Exposure time adjusted ______ Exposure time adjusted

underexposed

For Contrast Studies: The contrast must be placed over the detectors. If the abdomen, rather than contrast, is over the detectors, the image will be ____________.

decreases

Decreasing SID Exposure time ___________

increases

Decreasing kVp Exposure time/contrast ___________

Pathology report pathology -2; 1 25-30% QC

Density Adjustment: Allows the radiographer to adjust the amount of preset radiation detection values (for ___________ and not for patient size) Usually seen as ____, -1, 0, ____, 2 options Commonly initiates a _____% increment Routine use of these adjustments is an indication a service engineer should perform _____ calibrations

superimposed centered cover collimation

Detector Selection: Select the detector(s) that will be ________________ by the anatomy of greatest interest. The anatomy of interest must be ___________ properly over the selected detector(s). The patient's anatomy of interest must adequately _______ the detector combination. Improper ______________ may affect termination of the exposure time.

outside; center

Detector Selection: Typically, either the 2 ________ cells OR the _______ cell can be selected. Sometimes all 3

No Effect

Different types of image receptor _____ _______ on exposure time, but image quality may be affected

.50 sec .15 sec low; function

EXAMPLE: If the exposure time (manual technique) for a PA chest image is 0.20 sec., the backup time must be set longer than 0.20, such as 0.50 sec. The AEC will stop the exposure when a sufficient amount of radiation has reached the detector typically around 0.20 sec if AEC device fails, exposure will terminate at _____ sec. If the backup time was set only at 0.15 sec, the exposure will stop at _____ sec and the image will be underexposed Back up time operates under two conditions, if it is set too _____, or if AEC device does not __________ properly (operator error or mechanical error)

0.01 mA faster

Example: If the minimum reaction time is 0.01 sec., every exposure made with AEC needs to be at least _______ seconds long. Some times a problem for: -high ____ selection -pediatric -when ________ exposures are needed

increases

Excessive collimation Exposure time ___________

current directly small large timer terminated

How AEC functions: Exit radiation from the patient interacts with the detector, producing a _______ Strength of the current is ________ proportional to the exit radiation interacting with the detector So therefore, ______ current produced with a small amount of exit radiation ______ current produced with a large amount of exit radiation The current produced in the detector charges the xray _______ When the x ray timer receives a predetermined amount of charge from the detector, the x ray exposure is ___________

edge short; underexposure

If the _______ of a body part is positioned over the detector, the detector will detect some of the body part and some of the blank space. This will result in a _____ exposure and _____________.

increases

Increase patient thickness Exposure time ____________

increases

Increased density adjustment Exposure time ____________

increases

Increasing OID Exposure time ___________

decreases

Increasing mA Exposure time ___________

air electrical charge wire large

Ionization Chamber System: A hollow cell that contains ____ and is connected to the timer circuit by electrical wire. Ionization chamber is exposed, and the air inside the chamber becomes ionized, creating an _________ ________. Charge travels along the ______ to the timer circuit. The timer is tripped, and the radiographic exposure is terminated when a sufficiently ________ charge has been received.

terminate inherent underexposed 1 ms; (0.001)

The minimum reaction time is the shortest time it takes for the detector(s) to function and _________ the exposure. It is __________ in each system. If back up timer is set lower than the reaction time, AEC will not function properly and result in an image that is _______________ ____ ms in modern units (_______ s)

Photocell/Sensors/Detectors

The part of the AEC which detects when the desired amount of radiation has been received to achieve the desired IR brightness A set of three radiation-measuring detectors, arranged in some specific manner.

Ionization Chamber System

This represents which AEC system?

Phototimers

This represents which AEC system?

kVp mA detector arrangment density adjustment backup timer

When employing an AEC device, the radiographer will select the following settings on the control panel: (5)

5-6 NOT large

When the Backup time can be set: Many radiographers set the backup timer too high so that it is ___-___ times longer than the expected exposure time Do ______ set the backup time too high If AEC does not work, the patient will receive a _______ exposure

Outer

Which detectors do you activate for a AP Pelvis?

Center

Which detectors do you activate for a AP/Lateral Humerus?

Center

Which detectors do you activate for a Abdomen: Left Lateral Decub?

Ordinary

_________ detector arrangement

brightness mA

mA settings: Any mA can be selected for the image to produce an "optimum ___________" image, however... **the higher the ____, the lower the time

displayed feedback familiarity

mAs Readout: When AEC is employed, the total amount of mAs required to produce the IR exposure is determined by the system. This actual mAs amount is ____________ immediately after the exposure, sometimes just briefly. It is so important for the radiographer to take note of this readout -Can give immediate _________ on accuracy -Can help improve _____________ with exposure factors --Especially useful when imaging additional projections for the same patient


Conjuntos de estudio relacionados

Optometrist Assistant Stuff Part 2

View Set

Individual Health Insurance Policy Provisions 14

View Set

Interpersonal Comm Final (Ch. 1-4; 9-12)

View Set

AP Psychology Chapter 2: Research Methods Multiple Choice Part 2/2

View Set

Intro to Combustible Dust Hazards

View Set

Strategic Management Exam 1 Practice

View Set

Converting yards, feet, and inches

View Set