Chapter 2 Questions
If the depth is increased and the frame rate is unchanged, what else must have decreased?
# of lines per frame
Along with image depth, which of the following also determines the frame rate?
# of lines per frame, line density, # of focal zones, and imaging depth (PRF)
In an unfocused, single-element transducer, the width at the focal point of the sound beam measures how much compared to the beam width at the face of the transducer?
1/2
which of the following is the part of the transducer that stops the ringing of the element
Damping material
Which of the following would be considered the narrowest part of the sound beam?
Focal zone
which of the following electrical patterns produces electronic focusing of the ultrasound beam
curved
which of the following is true of the diameter of the sound beam in the fresnel zone( near zone)
decreases with distance from the transducer
along with crystal diameter, the divergence in the far field is also determined by what
frequency so frequency and crystal diameter determine divergence in far field
which of the following would cause an increase in frame rate
increased PRF decreasing the depth which is the same as increasing PRF, would increase the frame rate
Damping material produces all of the following except
increased SPL supposed to decrease the spa by decreasing # of cycles in a pulse
which of the following would cause a decrease in temporal resolution
increased line density
which of the following would most likely decrease beam divergence in the far field?
large crystal diameter, high frequency
which of the following would most likely increase the near zone length
large crystal diameter, high frequency
imaging transducers typically have
low Q factors, wide bandwidths
which resolution typically has the lowest # in clinical imaging
lower is better, axial is better than longitudinal, axial=LARD so answer is longitudinal
which of the following facilitates the transmission of sound from the element into the patients skin
matching layer
which of the following transducers can be described as having scan lines originate from a common point of origin
sector array comes to a point at the top, bc all of scan lines come from a point of origin
which type of resolution is an accurate representation of moving structures
temporal resolution, also known as frame rate, has ability to display moving structures in real time
which of the following is not a true statement about a mechanical transducer
there are no moving parts mechanical transducers DO have moving parts