Chapter 8 outline breakdown

¡Supera tus tareas y exámenes ahora con Quizwiz!

How many CI or chemical indicators should there be in a wrapped set of instruments

At least 2

Class 1

Autoclave tape

What type of spore is in dry heat

Bacillus

What type of spore is used in the BI for EO

Bacillus

When should a BI PCD test be performed diring a qualification testing of steam sterilization

Before the bowie dick test

Placement of BI in LTGP

Bottom shelf in the back perferablr in top if tray

Class 2

Bowie dick test pack

Didfernce between a CI and a BI

CI monitors the sterilization process BI contains active spores

Placement of BI and PCD in EO

Center of load

Placement of BI in LTHP

Center of the top shelf

What class of chemical indicator parallels to the use of a BI thus is used when sterilizing loads with implants

Class 5 integrating indicators

Parameters for ozone

Concentration temp time

Paramters for EO

Concentration time humidity temp pressure

How are microorganisims killed for steam

Condensation

What is a spore

Dehydrated gram-positive rods sell with a thick hard shell like covering capable of surviving extreme heat lack of water and exposure to toxic chemicals and radiation

Class 6

Emulating indicators

How to load Rigid sterilization containers

Flat on rack with 1inch space (2.5 cm)

What type of spore is used in LTGP

GEOBACILLUS

What type of spore is in ozone

Geobacillus

What type of spore is used in LTHP

Geobacillus

Class 3

Glass ampule

Function of the drain line

Heat sensing thermometer, coldest part of sterilizer which is the greatest area of challenfe for a sterilizer. Where the bowie dick and BI is placed for testing

Class 5

Integrating indicator

Define master product

Most difficult to sterilize tray from a family of products

Class 4

Multi variable indicator

How to load paper platoc pouches

On their edge side

How to load textiles

On thier side

What part of the steam sterilizer presents the greatest challange for sterilization

Over the drain. This is the best location to ensure an effective test

How are microorganisims killed for LTGP LTHP OZONE DRY HEAT

Oxidation

Placement of BI in ozone

Placed anywhere in the load

Function of the sterilizer/ insulated jacket

Provides heat inside the chamber to prevent condensation from forming on chamber walls help to separate water droplets from the steam before the steam is injected into the chamber

Bi test and color change for LTGP

Purple = negative Yellow=positive

Bi test and color change for LTHP

Purple =negative Yellow=positive

Bi test and color change for steam

Purple= negative Yellow= positive

Bi test and color change for EO

Purple= negative Yellow= positive

What is the first course of action is a bowie dick test fails for the first time

Re run the test and document

Defined superheating

Steam pressure is too low for set temptature When he is added to dry steam in the absence of water

What is gravity displacement sterilization

Steam using passive air displacement

What is dynamic air removal sterilization

Steam using pre vac sterilizers by condensation

What does sal stand for

Sterility assurance level

What does an incomplete color change of a class 2 indicator mean

Sterilizer does not have proper air removal theevacuation is not functioning properly l. The test must be reran

What part of the steam sterilizer surronds the internal chamber

Sterilizer jacket

Describe the BI PCD USED FOR OZONE STERILIZATION

Syringe in a pouch

Pacement of BI in dry heat

The fullest and densest load

Why is the presence of humidity critical to EO sterilization

The penetration of bacterial cells and successful sterilization

What is the meaning of a positive biological indicator

The sterilization process may not have been effective

How to load basin sets

Tilted on their side

How to load Solid trays

Tilted on thier sides

Parameters for dry heat

Time temp

Parameters for LTGP

Time temp hydrogen peroxide

Parameters for steam

Time temp pressure

Ehat is the purpose of a control BI

To cerify thay the spore is viable

The use of a bowie dick test

To test the air removal

Dwhat is the purpose for biologically testing rigid containers

To verify the sterilizer is able to sterilize the inside container

What is an ideal result for a bowoe dick test

Uniform color means proper air removal

Function of the inside/internal chamber of a sterilizer

Where the sterilization takes place where the steam is injected from the sterilizer jacket

What is the incubation temp of a BI for steam sterilization

131-140 degrees

What is the target temp anf time and pressure for dynamic air removal

270 @ 4 min 28-30 psi 275 @ 3 min 28-30 psi

Sterilizer setting for bowie dick test

3-3 1/2 minutes @270 no drying

How long should packs be allowed to cool

30-120 min

Parameters for sterilizer area

68° to 73° temp 30 to 60% humidity 10 Air changes positive pressure

What is the recommended release temperature for steam sterilizer packs

70- 80 degree

Parameters for sterilizer access area

75° to 85° 10 air changes negative pressure

How are microorganisims killed for EO

Alkylation

What is the package choice for dry heat sterilization

Aluminum foil


Conjuntos de estudio relacionados

HR: Working with Unions and Resolving Disputes

View Set

Anatomy & Physiology II Chapter 18 Guided Reading quiz

View Set

Econ 202-001 Prin Economics- Mirco

View Set

Comptia 220-801 4.6.6 Practice Test Questions

View Set

Durham - Ch 16 Discharge Planning and Teaching

View Set

AWS Certified Cloud Practitioner Study Guide 1

View Set

Spanish Affirmative and Negative Words (Opposites)

View Set

Personal Finance Chapter 6 Review

View Set

Pharmacology Ch. 16 Practice Questions

View Set

Chapter 12 - The Central Nervous System

View Set

España dos: La Peste y Monster Mash 98%=; 24/4

View Set