Dental Imaging Ch10
The coin test is used to check___ a)proper safelighting b)strength of the processing solution c)film density d)film-to-screen contact e)beam collimation
a)proper safelighting
How far away must a film be unwrapped from a safelight?
at least 4 feet
After processing, fresh film that has been properly stored and protected will appear ___. a) fogged b)clear with a light blue tint c)clouded with a blue tint d)dark blue e)totally black
b) clear with a light blue tint
One the average, processing solutions should be changed ___. a)once each day b)once each week c)every 3 to 4 weeks d)every 8 to 10 weeks e) every 3 to 4 months
c)every 3 to 4 weeks
Film processing required daily monitoring
...
When a fixer is at full strength, a film should clear within how many minutes?
2 minutes
The darkroom should be checked for safelighting every how many months?
6 months
After performing the film-screen contact test, a wire mesh image of uniform density appears. These results indicate ____. a)adequate film-screen contact b)inadequate film-screen contact
A) adequate film-screen contact
BACK OF THE CHAPTER QUESTIONS
BACK OF THE CHAPTER QUESTIONS
How should the strength of the developer be checked?
Compare film densities to a standard. -Reference radiograph -Stepwedge Radiograph -Normalizing device
Stepwedge
Composed of aluminum steps. Standard stepwedge radiograph View standard radiograph and daily radiograph Use the middle density for comparison If density differs by more than 2 steps, develope solution is depleted.
Basic elements of a quality administration program
Decription of the plan Assignment of duties Monitoring schedule Maintenance schedule Record-keeping log Plan for evaluation and revision In-service training
Who can perform annutal tests for dental x-ray machines?
Dentist Dental Hygienist Dental Assistant Manufacturer's service representative
What should cassettes be checked for?
Every month for worn closures, light leaks, and warping Adequate screen to film contact
Calibartion of dental x-ray equipment can be performed by a dentist, dental hygienist, or dental assistant. True or False
False, it must be done by a technician
What is the coin test and what does it check?
It checks the safelight. A coin is placed on an unwrapped film and the safelight is turned on from about 3-4 minutes. Then, the film is processed. If there is a visible image the safelight is not safe to use
What is the most critical component of film processing quality control?
Monitoring of the processing solutions
How often should screens be cleaned?
Monthly basis After cleaning the screen, an antistatic solution should be applied to it
Test to monitor fixer strength
Place an unwrapped film into fixer solution and measure the amount of time it takes for it to clear. If it takes 2 minutes, it is good. If it does not clear within 3-4 minutes the fixer is depleted.
What does a quality assurance plan include?
Quality control tests and quality administration procedures
How often should the processing solutions be replenished and how often should they be changed?
Replenished: daily Changed: Every 3-4 weeks
What is quality assurance?
Special procedures that are used to assure the production of high-quality diagnositic radiographs
What are quality control tests?
Specific tests taht are used to maintain and monitor dental x-ray equpiment, supplies, and film processing.
Annual tests for dental x-ray machines can be performed by a dentist, dental hygienist, or dental assistant. True or False
True
For quality control purposes, each new box of unopened film should be tested for film freshhness and fogging before it is used. True or False
True
Fresh films and fresh chemicals must be used when preparing reference radiographs True or false
True
When functioning properly, a viewbox should emit a uniform and brilliant light True or False
True
What is the test for film processing equpiment?
Unwrap 2 unexposed films and expose one to light. Process these films. One should come out clear and dry and the other should come out black and dry.
Reference Radiograph
Use fresh film and the correct exposure factors to make a reference radiograph. Chemicals must be at the recommended time and temperature. Compare this with radiographs taken daily.
What is a reference radiograph used to test?
Stregth of developer solution
Regardless of who actually exposes and processes the radiographs, the dentist is ultimately responsible for the diagnostic quality of all dental radiographs. True or False
TRUE
Who is responsible for overal quality assurance
The dentist
hat is quality administration
The management of the quality assurance plan in the dental office
The densities seen on the daily radiograph appear lighter than the densities seen on the reference radiograph. This result indicates that ___. a)The developer solution is too weark b)the developer solution is too concentrated c)the developer solution is too cold d)either a or c e)either b or c
d) either a or c
One the average, prcessing solutions should be replenished... (how often)
daily
The darkroom must be checked for light-tightness every how many months?
every month
The clearing test is used to monitor..
fixer strength
Normalizing device
monitors developer strength and film density
One of the most critical areas of quality control that requires daily monitoring is __. a)examination of the fluorescent bulbs inside the viewbox b)cleaning the extraoral intensifying screens c)examination of the darkroom for light-tightness d)processing of films e)all of the above
processing of films
What is the fresh film test and what do the results look like?
used to check a newly opened box of film. Process an unexposed film from a newly opened box. If the film is fresh it will appear to be clear with a slight blue tint. If the film is fogged it is expired, improperly stored, or exposed to radiation