Instrumentation, optimization and contrast- 6%

¡Supera tus tareas y exámenes ahora con Quizwiz!

When evaluating aortic regurgitation with CW Doppler in the apical view: · A. Set wall filter to the minimum level · B. Move the baseline to the top of the Doppler display · C. Set velocity scale so that the signal overfills the screen · D. Sweep speed set to 100mm/s

set sweep speed to 100mm/s

. Which of the following is usually better evaluated with Doppler than with a contrast exam? · A. thrombus formation in the left ventricular apex · B. shunt flow through a VSD · C. shunt flow through an ASD · D. persistent left SVC

shunt flow through vsd

Which imaging artifact is caused by non-specular reflections from a rough surface and wave interference? · A. speckle · B. multi-path · C. ghosting · D. misregistration

speckle

Which acoustic property causes the echogenic appearance of the pericardium related to surrounding structures? · A. refraction · B. non-specular reflection · C. specular reflection · D. attenuation

specular reflection

Which type of image resolution is negatively affected by the placement of multiple focal zones at the area of interest? · A. contrast · B. temporal · C. elevational · D. lateral

temporal

Which type of resolution improves with decreased image sector size? · A. temporal · B. contrast · C. elevational · D. axial

temporal

You suspect a calcification is present within the left ventricle. You activate Harmonics and add a focal zone to the field of view. Which type of resolution will degrade with these changes? · A. lateral · B. axial · C. temporal · D. contrast

temporal

The thickness of the interventricular septum will be best visualized when: · A. harmonics is turned off · B. the US beam is perpendicular to the long axis of the septum · C. the US beam is parallel to the long axis of the septum · D. the dynamic range is increased

the beam is perpendicular to the long axis of the septum

. A 27yr old patient presents for a follow up echocardiogram due to a history of a secundum ASD that was diagnosed two years ago. You are viewing the atrial septum in the subcostal view and apply color Doppler. No flow is detected across the septum. Which of the following could explain the lack of flow displayed across the defect? · A. The color Doppler map is inverted. · B. The color Doppler PRF is set too high. · C. PW Doppler is required to assess flow across the atrial septum because color Doppler is not well displayed in the subcostal view of the septum · D. Secundum ASDs commonly close spontaneously and the defect most likely has healed

the color doppler prf is set too high

The Doppler shift is: · A. the difference between the velocity of sound and the velocity of the blood flow · B. the amount of movement of blood flow that occurs each second · C. the difference between the speed of sound and the speed of the blood flow · D. the difference between the transmitted frequency and the reflected frequency.

the difference between the transmitted frequency and the reflected frequency

On an apical 4 chamber view, if you move the color Doppler sample box from the mitral level to the pulmonary veins, _______________. · A. the PRF increases · B. the sample size should be increased for better pulmonary venous flow evaluation · C. the frame rate decreases · D. temporal resolution will improve

the frame rate decreases

The pericardium can cause lobe artifact due to __________________________ · A. the high transmission rate of the beam through the tissue. · B. the close proximity to the diaphragm. · C. non-specular reflection of the smooth fibrous tissues. · D. the highly reflective nature of the tissue.

the highly reflective nature of the tissue

What effect will a mechanical index(MI) of 1.0 have on a contrast echo exam? · A. An MI setting of 1.0 or greater is preferred for contrast exams. · B. The increased beam intensity will burst the contrast bubbles quickly. · C. Mechanical index has no effect on contrast exams. The thermal index must be properly set for an optimal contrast exam. · D. The decreased beam intensity will reduce the appearance of the bubbles on the image.

the increased beam intensity will burst the contrast bubbles quickly

Increased overall gain levels on an ultrasound exam can cause ________________________. · A. underestimation of valve area on planimetry · B. increased posterior shadowing · C. myocardium to appear thinned · D. underestimation of Doppler velocity

underestimation of valve area on planimetry

How do you correct beam width artifact on a Doppler tracing? · A. use a smaller sample size · B. decrease the sweep speed · C. decrease the PRF · D. increase the incident angle of the Doppler cursor

use a smaller sample size

For optimal Doppler flow evaluation of the cardiac valves, a cursor angle of ____________ degrees should be used. · A. 0 · B. 45 · C. 60 · D. 90

0 degrees= parallel

All velocity and time interval measurements should be performed at a sweep speed of ___________ or greater. · A. 25 mm/s · B. 100 mm/s · C. 200 mm/s · D. 50 mm/s

100 mm/s

Frequencies used in transthoracic imaging are usually in the range of _______________. Frequencies used in transesophageal imaging are usually in the range of _______________. · A. 3-6MHz, 5-7MHz · B. 1-3MHz, 6-8MHz · C. 3-6MHz, 6-8MHz · D. 2.5-4MHz, 5-7MHz

2.504mhz, 5-7 mhz

A patient presents for an echocardiogram that is 6'3 and 320lbs. Which transducer would be the best transducer to use to perform the echo? · A. 4MHz · B. 5MHz · C. 2MHz · D. 3MHz

2mhz lower frequency=more penetration

Color scale settings should be set with a maximum velocity between ___________________cm/s for the routine echo exam. · A. 30 and 40 · B. 50 and 70 · C. 90 and 100 · D. 80 to 100

50-70

A 20yr old patient presents for an echocardiogram that is 5'3 and 110lbs. The exam is ordered to rule out sinus of Valsalva aneurysm. Which transducer would be the best transducer to use to perform the echo? · A. 4MHz · B. 1.75MHz · C. 5MHz · D. 2.5MHz

5mhz

The standard sector width for transthoracic echocardiography is: · A. 90 degrees · B. 180 degrees · C. 30 degrees · D. 60 degrees

60 degrees

. A Reynold's number ______________ indicates ________________. · A. >200, stenosis · B. >1.3, ventricular hypertrophy · C. >3.5, renal stenosis · D. >2000, turbulence

>2000, turbulence

Which of the following is best evaluated with perfluorocarbon contrast? · A. LVOT gradients · B. ASD · C. tricuspid regurgitation · D. VSD

LVOT gradients

Which of the following cardiac exam techniques provides the best spatial resolution of the cardiac structures? · A. resting echo with contrast · B. cardiac catheterization · C. transesophageal echo · D. stress echo

TEE

What is the major advantage of using color Doppler in transesophageal echo (TEE) evaluation of a prosthetic valve over color Doppler evaluation of the same valve on a transthoracic echo (TTE) exam? · A. TEE can identify paravalvular leakage which is usually very limited on a TTE. · B. TEE can identify turbulence caused by valvular stenosis which cannot successfully be diagnosed on TTE. · C. TEE can use color Doppler to assess the PISA radius of MR which is usually limited by valve masking on TTE. · D. TEE can identify valvular dehiscence while TTE usually cannot diagnose this abnormality with color Doppler.

TEE can identify paravalvular leakage which is usually very limited on a TTE

Which system control will simultaneously enhance far field echoes and diminish near field echoes? · A. overall gain · B. time gain compensation · C. rejection · D. threshold

TGC

If you increase the power output on the machine: · A. beam intensity and beam penetration increases · B. beam intensity and the magnitude of the Doppler shift increase · C. beam intensity, resolution and beam penetration increase · D. beam penetration and the magnitude of the Doppler shift increase

a beam intensity and beam penetration increases

Which of the following statements is true regarding the color flow evaluation of aortic stenosis? · A. The Reynold's number will be 1000-1500 in the area just distal to the stenosis. · B. When using a velocity map, aortic stenosis will cause a mosaic color pattern proximal to the stenosis. · C. A variance color map will demonstrate a green color in the aortic root, just distal to the stenosis · D. The greatest amount of turbulence seen with aortic stenosis is at the valve.

a variance color map will demonstrate a green color in the aortic root, just distal to the stenosis

Which of the following is an advantage of PW Doppler over CW Doppler? · A. ability to evaluate deeper flow patterns · B. not limited by PRF · C. no aliasing · D. ability to determine velocities at exact locations

ability to determine velocities at exact locations

While scanning in the apical 2 chamber view, you are trying to assess flow in a possible aneurysm in the thoracic aorta. Which of the following adjustments will assist in improving color flow in the aorta? · A. increase the color sample size · B. change the velocity color map to a variance map · C. activate the zoom function over the aorta · D. increase the wall filter setting

activate the zoom function over the aorta

While scanning an obese patient, you apply color Doppler to the mitral valve in the apical four chamber view. When you adjust the placement of the box, there is a delay between your adjustment and the box responding to the movement. How can you correct this? · A. switch to a higher frequency probe · B. activate the zoom function over the mitral valve · C. increase the sample size · D. decrease the color gain

activate the zoom function over the mitral valve

Which of the following would increase the frame rate? · A. increasing sector width · B. activating preprocessing zoom · C. adding a focal zone · D. increasing imaging depth

activating preprocessing zoom

Activating High PRF Doppler during evaluation of the LVOT velocity will cause the US system to: · A. reduce the Nyquist limit · B. switch to continuous wave Doppler · C. add a second PW Doppler sample volume · D. record the Doppler information faster

add a second PW Doppler sample volume

Harmonic imaging will: · A. allow for deeper penetration with improved lateral resolution · B. reduce the appearance of microbubble contrast on a contrast echo · C. degrade lateral resolution and improve axial resolution D. improve temporal resolution

allow for deeper penetration with improved lateral resolution

Contrast evaluation of the left heart can be used to differentiate: · A. aortopulmonary window from an outlet VSD · B. stunned endocardium from scar tissue · C. an LV aneurysm from an LV pseudoaneurysm · D. hibernating endocardium from scar tissue

an lv aneurysm from an lv pseudoaneurysm

At what point in the saline contrast echo will the patient perform the provocative maneuvers? · A. as soon as the contrast enters the right atrium · B. as soon as the contrast enters the left atrium · C. as soon as the contrast completely fills the right atrium · D. as soon as the contrast completely fills the left atrium

as soon as the contrast enters the right atrium

All of the following are advantages of contrast evaluation of shunt flow in the heart, except: · A. increases the strength of the Doppler signal displayed on the tracing · B. assists in proper placement of repair devices during corrective therapy · C. assists in proper placement of Doppler sample · D. aids in obtaining accurate velocity information across the defect

assists in proper placement of repair devices during corrective therapy

What console control does not change the amount or type of information displayed, only the perception of the images by viewer? · A. B-color · B. imaging depth · C. velocity scale · D. sector width

b-color

What acoustic artifacts can be mistaken for a vegetation on the native aortic valve? · A. side lobe · B. volume averaging · C. beam width · D. clutter

beam width

Which of the following statements best describes parabolic laminar flow? · A. Blood flow moves from an area of high pressure to low pressure. · B. Blood flow moves in layers that all move at the same speed. · C. Blood flow moves in layers with the highest velocity layer in the center of the vessel. · D. Small changes in vessel radius will cause significant changes in the resistance to flow.

blood flow moves in layers with the highest velocity layer in the center of the vessel

___________________ increases with increasing the overall gain. · A. The amount of attenuation · B. Image contrast · C. The amount of reflection · D. Brightness of the image

brightness of the image

Which of the following is a disadvantage of intracoronary ultrasound? · A. only allows for 90 degree imaging · B. cannot provide blood flow velocity information · C. high penetration but low resolution · D. can only be performed during open heart surgery

cannot provide blood flow velocity information

Which of the following correctly describes agitated saline contrast? · A. consists of a sodium chloride solution with bubbles small enough to easily pass through the pulmonary bed · B. consists of a nitrogen gas bubbles large enough to avoid passage through the pulmonary bed · C. consists of a sodium chloride solution with bubbles large enough to avoid passage through the pulmonary bed · D. consists of oxygen gas bubbles large enough to avoid passage through the pulmonary bed

consists of a sodium chloride solution with bubbles large enough to avoid passage through the pulmonary bed

Increasing the color Doppler sample size will __________________ the image frame rate. · A. decrease · B. increase · C. have no effect on D. only increase if you also decrease

decrease

Which of the following is related to degraded axial resolution on the image? · A. increase sector size · B. increase the number of focal zones · C. decrease in probe frequency · D. decrease imaging depth

decrease in probe frequency

Which of the following adjustments would improve the appearance of contrast on the echo image? · A. Use multiple focal zones · B. Turn off harmonic imaging · C. Increase the thermal index · D. Decrease output power

decrease output power

Which of the following adjustments will improve temporal resolution on color Doppler? · A. increase the wall filter · B. steer the sample box · C. increase color gain · D. decrease sample size

decrease sample size

Which of the following adjustments should be made to the Doppler settings when evaluating left ventricular inflow in a patient with constrictive pericarditis? · A. Decrease the sweep speed · B. Increase the sweep speed · C. Increase the PRF to demonstrate the expected E peak velocities greater than 2.0m/s · D. Decrease the PRF to demonstrate the expected low velocity peaks of the mitral waveform

decrease sweep speed

Increasing the imaging depth will: · A. decrease the pulse repetition frequency · B. increase the beam intensity · C. improve temporal resolution · D. decrease the pulse repetition period

decrease the PRF

While scanning the heart from the apical window, the ventricular chambers are anechoic but artifactual echoes appear to fill the atria. Which of the following adjustments of the system controls would best address this issue? · A. increase output power to increase penetration · B. decrease far field gain · C. decrease near field gain levels · D. switch to a higher frequency transducer

decrease the far field gain

. Increasing the imaging depth will: · A. decrease frame rate · B. increase the pulse repetition frequency · C. decrease the beam intensity · D. improve temporal resolution

decrease the frame rate

. You are performing an apical view during a microbubble contrast echo on a patient with suspected wall motion abnormalities in the basal anterolateral left ventricular segment. You notice significant swirling of the contrast in the area of the apex. How can you adjust the controls to reduce this artifactual swirling? · A. move the focal zone to the apex · B. turn off harmonic imaging · C. decrease the mechanical index setting · D. inject contrast at a faster rate

decrease the mechanical index setting

Which of the following adjustments will improve the temporal resolution on the image? · A. turning off the zoom function · B. turning on harmonic imaging · C. decreasing imaging depth · D. adding another focal zone

decreasing imaging depth

Which mode of US detects frequency shifts? · A. M mode · B. A mode · C. Doppler · D. B mode

doppler

Which of the following controls adjusts the number of shades of gray displayed of the 2D image? · A. output power · B. dynamic range · C. wall filter · D. monitor brightness

dynamic range

Which operator control can be used to adjust the image contrast? · A. dynamic range · B. overall gain · C. focal zone · D. output power

dynamic range

____________________ used to form the sector shaped field of view for a phased array transducer. · A. An acoustic lens is · B. Electronic steering is · C. Electronic focusing is · D. The curved arrangement of the elements are

electronic steering

Aliasing will occur when the Doppler shift ________________________. · A. exceeds twice the PRF · B. exceeds 1/2 PRF · C. is less than 1/2 PRF · D. is greater than the transducer frequency

exceeds 1/2 prf

Which of the following is a necessary machine setting when performing a contrast evaluation of the left ventricle to rule out thrombus? · A. Mechanical Index set to 1.0 or greater · B. Harmonic imaging off · C. Focal zone placement at the area of interest D. Low PRF color Doppler setting

focal zone placement at the area of interest

A high frame rate (≥30 frames per second) is preferred for cardiac imaging: · A. for accurate display of cardiac motion · B. because faster frames rates eliminate all acoustic artifacts · C. because faster frame rates improve penetration D. in order to reduce exam time and ultrasound exposure to the heart

for accurate display of cardiac motion

Harmonic imaging uses _______________________ to improve the image resolution by processing higher frequency reflections. · A. wave interference · B. frequency compounding · C. autocorrelation · D. apodization

frequency compounding

You are performing a bubble study to evaluate a patient with a possible pulmonary arteriovenous malformation(AVM). The standard echo demonstrates no shunt flow across the septum. Which of the following correctly describes what a positive study would demonstrate? · A. if bubbles are noted in the left atrium within three minutes of the injection, a pulmonary AVM is present · B. if bubbles are noted in the right atrium within four to eight cardiac cycles after the injection, a pulmonary AVM is present · C. if bubbles are noted in the left atrium within four to eight cardiac cycles after the injection, a pulmonary AVM is present · D. if bubbles are noted in the left atrium within 60 seconds of the injection, a pulmonary AVM is present

if bubbles are noted in the left atrium within 4-8 cycles after injection, a pulmonary avm is present

. Why is the temperature of the TEE probe monitored? · A. If the PE element is heated over 40 degrees Celsius it would cause the plastic housing to melt · B. If the PE element is heated over 40 degrees Celsius it would cause the element to lose its ability to produce soundwaves · C. If the PE element is heated over 40 degrees Celsius it would cause damage to the tissues of the esophagus/stomach · D. If the PE element is heated over 40 degrees Celsius it would cause the element to decrease the frequency of the soundwaves produced

if over 40 degrees celsius it could damage the esophagus/stomach

Why can't transesophageal probes be thermally sterilized? · A. because they do not detach from the ultrasound machine · B. if the probe is heated above the Curie temperature, the piezoelectric properties will be lost · C. because the electrical wiring cannot get wet · D. because the plastic components would probably melt

if the probe is heated above the curie temperature, the piezoelectric properties will be lost

Which of the following system settings should be kept the same during resting and stress images for a stress echo? · A. focal zone placement · B. overall gain · C. compression · D. image depth

image depth

Increasing transducer frequency will have what effect on scatter of the ultrasound beam? · A. no change · B. decrease · C. unable to answer without knowing how the elements are arranged and fired · D. increase

increase

Increasing the transducer frequency will: · A. improve temporal resolution · B. improve axial resolution but degrade lateral resolution · C. increase the Doppler shift detected D. decrease beam intensity

increase the doppler shift detected

Which of the following will improve the lateral resolution on the image? · A. decrease the PRF · B. adjust dynamic range · C. increase the focal zone number D. decrease transducer frequency

increase the focal zone number

While performing a Doppler evaluation of the aortic valve, a significant amount of wall motion artifact is displayed on the Doppler tracing. How can you eliminate the appearance of the artifactual information on the Doppler tracing while maintaining accurate display of the aortic waveform? · A. decrease the velocity scale · B. decrease the wall filter settings · C. decrease the Doppler gain · D. increase the high pass filter settings

increase the high pass filter settings

Which of the following will aid in visualization of the endocardium? · A. increased transducer frequency · B. larger sector size · C. decreased PRF · D. agitated saline contrast

increase the transducer frequency

The ____________________ of the signal displayed on CW Doppler is directly related to the severity of the regurgitation. · A. peak velocity · B. peak pressure gradient · C. size of the spectral window · D. intensity

intensity

The Doppler shift _____________________________ the blood flow velocity. · A. is always higher than · B. is always lower than · C. is directly proportional to · D. is indirectly proportional to

is directly proportional to

You suspect a calcification is present within the left ventricle. You activate Harmonics and add a focal zone to the field of view. Which type of resolution improves the most? · A. axial · B. contrast · C. elevational · D. lateral

lateral

Which of the following technical factors can negatively affect the 2D planimetry evaluation of the mitral valve? · A. transducer with a short spatial pulse length and high frequency · B. transducer with a short spatial pulse length and long wavelength · C. MVR valve masking and lateral resolution · D. lateral resolution and gain settings

lateral resolution and gain settings

Which of the following ultrasound techniques has the highest temporal resolution? · A. 2D imaging · B. 3D imaging · C. M-mode · D. Color Doppler

m-mode

______________ refers to shadowing caused by mechanical prosthetic valvular apparatus that inhibits flow display on a color Doppler image. · A. Ghosting · B. Masking · C. Reverberation · D. Clutter

masking

The coupling gel used for ultrasound exams provides the same function as what transducer component? · A. PE element · B. matching layer · C. backing layer · D. damping layer

matching layer

The ultrasound machine ___________________ the Doppler shift and ________________ the Doppler velocity. · A. measures, estimates · B. calculates, measures · C. measures, calculates · D. produces, evaluates

measures,calculates

Adjustment of which of the following imaging controls would be best to assist in improving visualization of the descending aorta posterior to the left atrium on an obese patient? · A. increase probe frequency · B. move focal zone to the level of the aorta · C. increase output power · D. use saline contrast injection

move the focal zone to the level of the aorta

. _________________ adjusts the brightness of the entire image, while ________________ adjusts the brightness of specific portions of the image. · A. monitor contrast, time-gain compensation · B. overall gain, output power · C. overall gain, time-gain compensation · D. post-processing, pre-processing

overall gain, TGC

Overgaining on an ultrasound exam can cause ________________________. · A. overestimation of valve area on planimetry · B. blood to appear anechoic · C. overestimation of Doppler velocity · D. increased posterior shadowing

overestimation of doppler velocity

Which of the following is an imaging technique that is not dependent on the angle of incidence of the sound beam? · A. Power Doppler · B. Speckle tracking · C. Color Doppler · D. CW Doppler

power doppler

. While scanning an echo, you change the gray map to a new choice and continue to scan. This would be termed a change in _____________________. · A. harmonic imaging · B. post-processing · C. pre-processing · D. bandwidth

pre processing

What type of magnification provides the highest frame rate and best image resolution? · A. complete · B. preprocessing · C. partial · D. postprocessing

preprocessing

The ultrasound system relies on the __________________________ for accurate recording of motion clips on an echo. · A. system timer to be set at 3 seconds · B. system timer to be set at 5 seconds · C. R wave and T wave on the EKG · D. QRS wave on the EKG

r wave and t wave on the ekg

Which of the following is a non-specular reflector that will scatter the sound waves? · A. The endocardium · B. The pericardium · C. The myocardium · D. Red blood cells

red blood cells

Which of the following describes a purpose for applying acoustic gel to the face of the transducer? · A. To increase the amount of reflection that occurs within the first 2cm of tissue · B. To reduce the beam intensity just before the sound beam enters the patient · C. To reduce the difference in impedance between the matching layer and the skin · D. To increase the amount of refraction that occurs within the first 2cm of tissue

reduce the difference in impedance between the matching layer and the skin

Which of the following refers to a dedicated CW (pedoff) probe? · A. Can only be used at the suprasternal window · B. Preferred for documentation of the location of the highest velocity · C. Requires correlation with the EKG tracing to confirm where the waveform is coming from · D. Single piezoelectric element

requires correlation with the ekg tracing to confirm where the waveform is coming from

Transesophageal transducers are considered _____________ devices that should be disinfected using _________________. · A. critical, high level disinfectant · B. semi-critical, thermal sterilization · C. semi-critical, high level disinfectant · D. critical, thermal sterilization

semi-critical, high level disinfectant


Conjuntos de estudio relacionados

LESSON 5 zhè shì wǒ de jiā 这是我的家 (This Is My Family)

View Set

Shock, Sepsis & Multiple Organ Dysfunction NCLEX

View Set

***HURST REVIEW NCLEX-RN Readiness Exam 2***

View Set

Lesson 5: Public Restrictions on Land Quiz

View Set

Basic Learning Concepts and Classical Conditioning

View Set