IAHCSMM Central Service - Chapter 14 High-Temp Sterilization

Réussis tes devoirs et examens dès maintenant avec Quizwiz!

Factors that impact sterilization effectiveness

-The type of microorganisms present -The design of the medical device -The number of microorganisms (bioburden) present -The amount and type of soil present; soil acts as a shield for microorganisms

Cooling time depends on

-Type of sterilizer used -Design of the device and packaging being sterilized -Temp and humidity of the room

Steam-Flush Pressure Pulse Sterilizers

-Use a repeated sequence of a steam flush and pressure pulse to remove air from the sterilizing chamber and processed materials -Air removal occurs above atmospheric pressure; no vacuum is required

Conditions Necessary for Effective Steam Sterilization

Contact Temperature Time Moisture

Temperature Condition

Gravity sterilization - 250F Dynamic Air removal - 270F - 275F

Bioburden

The number of microorganisms on a contaminated object; also called "bioload" or "microbial load"

Saturated Steam

Steam that contains the maximum amount of water vapor

Jacket

-CS departments typically use jacketed sterilizers. -The interior chamber walls of the sterilizer are heated by steam in the metal jacket; helps minimize the amount of condensation that forms when hot steam contacts the chamber -Jacket surrounds the sides, top, bottom of the vessels and steam circulates in this space to preheat the interior chamber walls -Outside of jacket is covered with insulation

Phases of a Saturated Steam Sterilization Cycl

-Conditioning - steam enters at the upper back portion of the sterilizer; air is displaced through the drain; pressure begins to rise, as does the steam temperature Exposure - After the desired temperature is reached, the sterilizer's control system begins timing the cycle's exposure phase Exhaust - After exposure phase, drain is opened and steam is removed through the discharge line; filtered air gradually reenters the chamber and the chamber returns to room pressure Drying - drying times are based on the device, packaging and the sterilizer's IFU

Tabletop Sterilizers

-Frequently used in clinics and dental offices -Use electricity to heat water into steam

Immediate Use Steam Sterilizers

-Located in ORs or surgical suite substerile rooms, L&D units and special procedure areas that perform invasive procedures; intended for emergencies, have little to no drying time

Thermostatic Trap

-Located in the drain line -The drain and the area surrounding it are the coolest areas in the sterilizer -A sensor in the chamber drain measures steam temp and automatically controls the flow of air and condensate from the sterilizing chamber

Advantages of Steam Sterilization

-Low cost -Rapid sterilization cycles -Relatively simple technology -Leaves no chemical residues or byproducts

Contact Condition

-Most common reason for failure Caused by: -Inadequate cleaning -Sets are too dense -Poor instrument positioning -Packages wrapped too tight -Loads are too crowded -Basin positioning -Clogged drain strainer -Mechanical malfunctions -Utility malfunctions

Causes of Wet Packs

-Packs were improperly prepared or loaded -heavy or dense instrument sets -Not using absorbent material to wick moisture between heavy metal (basins) -Textiles wrapped too tight -Items wrapped while still damp or moist -Gasket not intact -Clogged chamber drain screen etc.

Gauges and Controls

-Provide a visual and written record of sterilization conditions Printout contains : -Date and time the cycle began -Selected cycle parameters, such as type of cycle, sterilization temp and dry times -written record of actual cycle activities (temp, exposure times, pressure)

Dynamic Air Removal

-Similar to gravity air removal, except there is a vacuum pump or water ejector. -Usually operate at higher temperatures (270F-275F) than gravity. -Preconditioning phase increases the speed of operation and reduces the chance of air pockets. -Uses variations of prevacuum air removal and above-atmospheric-pressure processes, such as steam-flush pressure-pulse process (SFPP) -Effective air removal is critical for steam penetration

Gravity Air Displacement

-Small to medium-sized sterilizers have this. -Air is heavier than steam, the steam forces the cooler air to the bottom of the chamber and out the drain.

Prevacuum Steam Sterilizers

-The dynamic air removal cycle depends on one or more pressure and vacuum sequences at the beginning of the cycle to remove air during the preconditioning phase -Typical operating temperatures are 270F - 275F -Bowie Dick test should be processed daily to ensure proper air removal

Door, Gasket, and Chamber Drain

Door - weakest part of a steam sterilizer; has safety locking mechanism when pressurized and can only be unlocked after depressurizing Door Gasket- designed to maintain a tight seal to prevent steam from escaping the chamber and air entering the chamber Chamber Drain - located on the front center of the floor -Must be cleaned daily, debris can impede performance

Moisture

Dry, saturated steam is required for effective sterilization -Relative humidity should be 97% - 100% -Pressure must be increased to increase steam temperature

Time Condition

items within the load must be exposed to high temperatures for an adequate amount of time


Ensembles d'études connexes

Financial Accounting Midterm review

View Set

патфіз тести модуль 1

View Set

AP Government Chapter 4: Civil Liberties

View Set

Introduction to Philosophy FINAL!!!!!!!

View Set

Ch 8 commercial Property Insurance

View Set

Organic Chemistry Distillation Quiz (Lab 1)

View Set

Gen Bio II: Ch. 35 Connect LS questions

View Set

Chapter 11 Sales "Making the presentation"

View Set