36. Smaller transducers always produce smaller beam diameters a) True b) False

b) False Beam diameter depends on whether of not the transducer is focused, the distance from the transducer, and whether one is in the near field or far field. Small transducers produce greatly diverging beams in the far field

7. Grating lobes are essential for the proper operation of a linear phased array. a) True b) False

b) False Grating lobes are an undesirable property of multi-element array transducers

39. The gray-scale display was made possible when digital scan converters replaced analong scan converters a) True b) False

b) False Gray-scale images were also possible with analog scan storage devices

40. Videotape recorders operating at 30 frames per second cannot be used with mechanical real-time units because the frame rate is too slow a) True b) False

b) False If the imaging rate is slower than the video framing rate, frames are duplicated to fill in

44. The near zone length should be kept as short as possible a) True b) False

b) False Images produced in the near zone have the best lateral resolution; therefore, the near zone length should be long enough to include all time regions-of-interest

41. Linear phased array scanners need no form of mechanical focusing because focusing is performed electronically a) True b) False

b) False Linear arrays must also be focused in the direction perpendicular to the array axis to reduce the slice thickness. This is usually accomplished by mechanical focusing

35. No refraction can occur at an interface if the media impedances are equal a) True b) False

b) False Refraction comes as a result of changes in sound velocity across a boundary, not impedance

8. Electronically steered scanners always produce higher resolution images than do mechanically steered scanners. a) True b) False

b) False Resolution is a function of the focusing characteristics of any scanner and cannot be assumed to be better for any particular scanner configuration

27. In a pulse-echo system, the quality factor should be made as large as possible a) True b) False

b) False The Q factor (quality factor) of a transducer refers to the length of time that the sound persists. High-Q transducers tend to ring for a long time, whereas low-Q transducers ring for a shorter time. A low Q produces a short pulse length, which is necessary to provide good images

28. The beam diameter is constant in the near zone a) True b) False

b) False The beam diameter reduces to 1/2 of the transducer diameter from the face of the transducer to the focal point

10. A linear sequenced array cannot by dynamically focused a) True b) False

b) False The focusing of a linear array is controlled by timing and is thus capable of dynamic function

54. Receiver-angle values in CW Doppler analysis do NOT influence the final Doppler shift a) True b) False

b) False The magnitude of the Doppler shift frequency is directly realted to the cosine of the angle between the direction of the blood flow and the receiver angle. The maximum shift frequency is, therefore, observed when the flow is along the receiver line of sight

67. Gray-scale systems typically use ______ as a means of signal dynamic range reduction a) rejection b) compression c) relaxation d) elimination

b) compression

4. Lateral resolution a) and ring-down are the same b) depends on the beam diameter c) improves with frequency d) cannot be measured in the far zone

b) depends on the beam diameter Lateral resolution (LATA: lateral, angular, transverse, azimuthal resolution) is defined as being equal to the beam diameter

56. Reflection factor at an interface between two materials depends primarily on the change in ______ across the interface a) acoustic impedance b) wave velocity c) density d) none of the above

a) acoustic impedance

20. The Doppler shift frequency is a) directly proportional to the velocity of the reflector b) greater in pulsed Doppler systems c) greater at high intensity levels d) dependent on the number of transducer elements being used

a) directly proportional to the velocity of the reflector the higher the velocity, the higher the shift (the higher the Nyquist limit)

71. The fraction of time that pulsed ultrasound is actually on is the a) duty factor b) frame rate c) cavitation d) intensity transmission coefficient

a) duty factor

46. Axial resolution can be improved by using a) higher frequency transducers b) lower frequency transducers c) larger transducers d) poorly damped transducers

a) higher frequency transducers Higher frequency transducers usually produce shorter spatial pulse lengths and thus improve axial resolution

25. The attenuation for soft tissue is a) increased with tissue thickness b) determined by the scope of the TCG curve c) increased with decreasing wavelength d) unimportant when using digital scan converters

a) increased with tissue thickness Attenuation is the product of the attenuation coefficient and the path length

47. When particle motion of a medium is parallel to the direction of a wave propagation, the wave being transmitted is called a a) longitudinal wave b) shear wave c) surface wave d) lamb wave

a) longitudinal wave

21. The number of frames per second necessary for a real-time image to be flicker free is a) more than 15 b) less than 10 c) between 6 and 10 d) between 3 and 6

a) more than 15

59. The change in direction of an ultrasonic beam, when it passes from one medium to another, in which elasticity and density differ from those of the first medium is called a) refraction b) rarefaction c) angulation d) reflection

a) refraction

63. What is the following formula used to determine? (2 x velocity of reflector x original frequency)/(velocity of sound) a) shift in the frequency caused by the Doppler effect b) degree in attenuation c) distance a wave travels d) amount of amplification necessary to produce diagnostic ultrasound

a) shift in frequency caused by the Doppler effect Doppler shift equation

60. A long near zone can be obtained by a) using a higher frequency transducer b) adding a convex lens to the transducer c) decreasing the diameter of the transducer d) increasing the damping

a) using a higher frequency transducer NZL= (ap)^2/4 x wavelength

29. The operating frequency a) depends on the transducer's ring-down time b) depends on the thickness of the crystal c) is increased as the crystal diameter is decreased d) depends on the strength of the pulser

b) depends on the thickness of the crystal Thickness equals 1/2 the wavelength

70. Most of the scan converters currently in use are of the ________ type. a) analog b) digital c) bistable d) static

b) digital

62. The ratio of particle pressure to particle velocity at a given point within the ultrasonic field is a) interference b) impedance c) incidence d) noise

b) impedance Z= particle pressure/ particle velocity Z= density x c

18. Propagation speed errors result in a) reverberation b) improper axial position c) shadowing d) a Doppler shift

b) improper axial position Improper axial resolution (along the beam) position

33. Increasing the pulse repetition period a) improves resolution b) increases the maximum depth that can be imaged c) decreases the maximum depth that can be imaged d) increases refraction

b) increases maximum depth that can be imaged

74. The major factor in determining the acoustic power output of the transducer is the a) size of the transducer b) magnitude of the voltage signals c) amount of amplification at the receiver d) amount of gain

b) magnitude of the voltage signals magnitude of voltage spikes applied to the transducer by the pulser

73. Which of the following image shapes is produced by a linear sequenced array real-time transducer? a) trapezoidal b) rectangular c) circular d) diamond

b) rectangular linear sequenced= rectangular

79. The substance typically used as a matrix on the imaging surface of an analog scan converter tube is a) mercurous chloride b) silicon oxide c) manganese sulfide d) lithium fluoride

b) silicon oxide Silicon oxide matrix is charged by a scanning electron beam, producing an image pattern that can be displayed on a video monitor

65. A 10-dB difference in signal intensity is equivalent to a ______ difference a) twofold b) tenfold c) hundredfold d) thousandfold

b) tenfold 3 dB- 2x 6 dB- 4x 9 dB- 8x 10 dB- 10x 20 dB- 100x

23. The intensity of the ultrasound beam is usually greater at the focal zone because a) decreased attenuation b) the smaller beam diameter c) diffraction effects d) a shorter duty factor

b) the smaller beam diameter intensity is defined as power per unit beam area; as the beam area decreases the intensity increases

66. The level below which signals are not transmitted through an ultrasound receiver system is the a) intensity level b) threshold, negative, or reject level c) impedance level d) dynamic range level

b) threshold, negative, or reject level

2. The spatial pulse length a) determines penetration depth b) usually decreases with frequency c) is improved with rectification d) determines lateral resolution

b) usually decreases with frequency SPL is defined as the product of the number of cycles in the pulse and its wavelength. This is generally shorter for higher frequencies since the wavelength is shorter

55. Refraction occurs because of difference in ________ across an interface between two materials a) acoustic impedance b) wave velocity c) density d) none of the above

b) wave velocity Refraction is described by Snell's Law, which relates that the incident angle to the transmitted angle, to the relative velocities of the two media making up the interface

12. Axial resolution

Row A

13. Lateral resolution

Row B

14. Dead zone

Row D

11. Registration or B-mode alignment

Rows A, B, C, D, and E All rods must be used to check registration accuracy

15. Depth calibration

Rows C or E

50. The equation that describes the relationship among propagation, wavelength, and frequency is a) C= frequency x wavelength b) wavelength= 2(frequency x C) c) Z= density x C d) wavelength= frequency + C

a) C= frequency x wavelength

48. The wavelength in a material having a wave velocity of 1500 m/s employing a transducer frequency of 5MHz is a) 0.3 mm b) 0.3 cm c) 0.6 mm d) 0.6 cm

a) 0.3 mm (1.5 mm/us)/ 5MHz = 0.3 mm

61. If a 2MHz frequency is used in human soft tissue, the wavelength is approximately a) 0.75 mm b) 0.15mm c) 0.21 mm d) 0.44mm

a) 0.75 mm

68. The strength of the echo is related to the height of deflection on the oscilloscope for the _______ display a) A-mode b) B-mode c) B-scan d) M-mode

a) A-mode

26. The acoustic impedance of the matching layer a) can be chosen to improve transmission into the body b) must be much larger than the transducer to reduce attenuation c) is not necessary with real-time scanners d) must be made with the same material as the damping layer

a) Can be chosen to improve transmission into the body Should be chosen to be at a value approximately equal to the mean of impedances of the material on the other side of it

37. The ultrasound beam profile can be measured with a hydrophone a) True b) False

a) True A hydrophone is usually a very small piezoelectric crystal that can be moved in front of the transmitting crystal to map its beam characteristics

9. An annular array scanner uses mechanical beam steering a) True b) False

a) True Annular arrays may be used for creating variable focusing in two dimensions but are not capable of electronic steering of the beam.

43. It is not possible to steer an annular phased array electronically a) True b) False

a) True Annular phased arrays must be steered mechanically, most frequently through the use of an oscillating acoustic mirror.

38. The pulsed-Doppler system yields better depth resolution than the continuous wave (CW) system. a) True b) False

a) True Pulse Doppler utilizes range gating to localize the signals to specific depths

The intensity of an ultrasonic beam can be measured with radiation force balance a) True b) False

a) True Radiation pressure can be measured with a sensitive balance and converted to intensity

42. The frame rate of real-time scanners depends on the number of lines used to form the image a) True b) False

a) True Real-time scanners display 15-60 images per second, called the frame rate. This rate must be greater than 15 frames per second to produce flicker-free images. Each frame is made of scan lines per frame, and the number of frames per second are related to one another.

22. The SPTA intensity will always be larger than the SATA intensity a) True b) False

a) True Spatial Peak (SP) values will always be greater than Spatial Average (SA) intensity values

17. How much will a 3.5 MHz pulse be attenuated after passing through 2cm of soft tissue? a) 7 dB b) 3.5 dB c) 17 dB d) 1.75 dB

b) 3.5 dB A rule of thumb approximating the attenuation coefficient of a reflected echo in soft tissue is 0.5 dB/cm/MHz. This, the attenuation coefficient will be 1/2 of the operating frequency. Attenuation= (A.C. dB/cm) x (path length cm) dB= 1.75 dB/cm x 2cm= 3.5 dB

53. The average velocity of ultrasonic waves in soft biological tissue is a) 1540 ft/s b) 3300 m/s c) 1540 m/s d) 300 x 10^6 m/s

c) 1540 m/s

45. What percentage of intensity of an ultrasound pulse incident on an interface of 0.25 and 0.75 rayls is reflected? a) 50% b) 100% c) 24% d) 75%

c) 25% Reflection coefficient= {(Z2-Z1)/(Z2+Z1)} squared x 100 {(0.75-0.25)/(0.75+0.25)}^2 x 100= 25%

64. The principle that states that all points on a wavefront can be considered as point sources for the production of spherical secondary wavelets was postulated by a) Doppler b) Young c) Huygens d) Langevin

c) Huygens

30. The period of an ultrasound wave is a) the time at whichit is no longer detectable b) determined by the duty factor c) the time of one wavelength d) independent of the frequency

c) time of one wavelength

1. An example of a nonspecular reflector is a) the liver surface b) the diaphragm c) red blood cells d) any structure that does NOT produce a strong echo

c) Red blood cells Structures smaller than a wavelength will scatter sound in all directions and are sometimes called diffuse reflectors

34. Ultrasound bioeffects a) do not occur b) do not occur with diagnostic instruments c) are not confirmed below 100 mW/cm2 SPTA d) are not confirmed below 1 W/cm2 SPTA

c) are not confirmed below mW/cm2 SPTA

52. The acoustic impedance of a material is a) directly proportional to density and inversely proportional to velocity b) directly proportional to velocity and inversely proportional to density c) inversely proportional to density and velocity d) equal to the product of density and velocity

c) directly proportional to density and velocity

16. Decreasing the spatial pulse length a) reduces the field of view b) reduces the lateral resolution c) improves axial resolution d) improves lateral resolution

c) improves axial resolution Decreasing the SPL improves the axial resolution. Axial resolution is equal to 1/2 of the SPL (LARRD=SPL/2)

5. The beam of an unfocused transducer diverges a) because of inadequate damping b) in the Fresnel zone c) in the Fraunhofer zone d)when the pulse length is long

c) in the Fraunhofer zone The beam of an unfocused transducer diverges in the Fraunhofer zone (far zone)

24. If the amplitude is doubled, the intensity is a) doubled b) cut in half c) increased by 4 times d) unchanged

c) increased by 4 times Intensity equals the square of the amplitude over the area

58. Increasing the frequency of an ultrasonic longitudinal wave will result in _____ in the velocity of that wave a) an increase b) a decrease c) no change d) a reversal

c) no change

76. The _______ relates bandwidth to operating frequency a) near zone b) piezoelectric crystal c) quality factor d) far zone

c) quality factor (Q factor) Q= operating frequency/ bandwidth

77. The extraneous beams of ultrasound generated from the edges of individual transducer elements and not in the direction of the main ultrasonic beam are called a) phased array b) impedance artifacts c) side lobes d) acoustic errors

c) side lobes

57. The velocity of sound waves is primarily dependent on a) angulation b) reflection c) the material through which the sound is being transmitted and the mode of vibration d) none of the above

c) the material through which the sound is being transmitted and the mode of vibration

6. Reverberation artifacts are a result of a) electronic noise b) improper TGC settings c) the presence of two or more strong reflecting surfaces d) an angle of incidence that is too small

c) the presence of two or more strong reflection surfaces Reverberation artifacts are present when two or more strong reflectors are located within the beam. One of these may be the transducer itself. (ping pong effect)

80. Specular reflection occurs when a) the frequency is small compared with the wavelength b) the object that causes the reflection is small c) the reflector surface is smooth as compared with the wavelength d) the angle of incidence and the angle of reflection differ by at least 45 degrees

c) the reflector surface is smooth as compared with the wavelength Reflectors whose boundaries are smooth relative to a wavelength behave as mirrors and are called specular reflectors

51. Which of the following can occur when an ultrasonic beam reaches the interface of two dissimilar materials? a) reflection b) refraction c) mode conversion d) all of the above

d) all of the above Reflection refers to echoes; refraction refers to the bending of the beam; and mode conversion refers to a change in the propagation mode transverse, shar, surface, or longitudinal waves.

3. Axial resolution a) is improved in the focal zone b) depends on the TGC slope c) is improved by the digital scan converters d) depends of the wavelength

d) depends on the wavelength Axial resolution, also called longitudinal, range, radial, or depth resolution (LARRD), is determined by the wavelength damping and frequency. Axial resolution improves with increased frequency

31. The dynamic range of a system a) is increased when specular reflectors are scanned b) is decreased when shadowing is present c) can be increased through the use of coupling gel d) is the ratio of smallest to largest power levels that the system can handle

d) is the ratio of smallest to largest power levels that the system can handle

78. The product of the period and the number of cycles in the pulse is the a) pulse repetition frequency b) continuous wave c) pulse repetition period d) pulse duration

d) pulse duration

69. Which of the following is NOT a method for restricting the dynamic range of the signal? a) suppression b) rejection c) compression d) relaxation

d) relaxation Relaxation processes are modes by which ultrasound may be attenuated in passing through a material. Suppression is another word for rejection

72. The ability of an imaging system to detect weak reflections is called a) compression b) demodulation c) gain d) sensitivity

d) sensitivity

49. The factor that determines the amount of reflection at the interface of two dissimilar materials is a) the index of refraction b) the frequency of the ultrasonic wave c) Young's modulus d) difference in specific acoustic impedances

d) the difference in specific acoustic impedances The fraction of sound reflected at an interface (r) is given by: R= {(Z2-Z1)/(Z2+Z1)} squared x 100

75. Two primary mechanisms that produce tissue biological effects are a) oscillations and radiation b) absorption and reflection c) direct and indirect d) thermal and cavitational

d) thermal and cavitational

19. Enhancement is caused by a) strongly reflected structures b) propagation speed errors c) Snell's Law d) weakly attenuating structures

d) weakly attenuating structures ex) think of the bladder (attenuation rate is slower through fluid than SBT, so enhancement is seen posteriorly to bladder


Ensembles d'études connexes

DECA Hospitality and Tourism Cluster Study Set

View Set

Chapter 8: Early Jewish, Early Christian and Byzantine Art

View Set

FFA Dairy Foods Competition-Cheese Identification

View Set

Tropical Weather & Forecasting Final Exam

View Set

Chapter 12: International Bond Market

View Set

Chapter 7, 8, 9, and 10 Astronomy test

View Set