CH 29 Quality assurance and control

अब Quizwiz के साथ अपने होमवर्क और परीक्षाओं को एस करें!

IR ESD

1,000 mGy (100rad)

To assess Specular reflectance turn off the monitor and look for illumination within a _ angle of viewing at approximately _ to _ cm.

15 degree, 30, 50

Image Display (monitors): Dead or Stuck Pixels

15(entire screen) 3(1 cm circle) 3 (adjacent) Naked eye or Magnifying glass

ACR and AAPM recommend a minimum resolution of electronically displayed images of ?

2.5 Lp/mm

Image Display (monitors): Display Resolution

2.5 Lp/mm (white/black difference) TG18-CX pattern

A 4 min fluoroscopic exam with a 1.5mA tube current results in a patient dose of _ rad?

2xminxmA = 2x4x1.5 = 12 rad

Luminance uniformity should not deviate more than _ from their average?

30 %

ESD (Entrance Skin Dose)

30 - 50 mGy/min (3-5 R/min) Shall not exceed 100 mGy/min (10R/min)

Mobile C-arm fluoroscopy units have an SSD of ?

30 cm (12 inches)

Stationary Fluoroscopic units have a source to skin distance (SSD) of _?

38 cm (15 inches)

AAMP

American Association of Physicists in Medicine

ACR

American College of Radiology

Quality Control: AEC Exposure Reproducibility

Annual +/- 1% (OD) Densitometer (phantom)

Quality Control: AEC Ion Chamber (IC) Sensitivity

Annual +/- 10 Exposure meter

Quality Control: AEC Response Capability

Annual +/- 10 % Exposure meter (lucite phantom)

Quality Control: Exposure Linearity

Annual +/- 10% Dosimeter

Quality Control: Exposure time

Annual +/- 5% Exposure meter or spinning top

Quality Control: kVp

Annual +/- 5% kVp +/- 10% kVp Bushong KVp meter

Fluoroscopic QC: Exposure Rate

Annual < 10 rad/min Exposure meter

Fluoroscopic QC: Protective Apparel

Annual No gaps or cracks Fluoroscope or film/IR @120 kVp

Fluoroscopic QC: ABC (AERC)

Annual No limits (check different attenuators) Exposure meter

Fluoroscopic QC: Resolution

Annual No limits (look for higher Lp/mm) Resolution phantom

Quality Control: Filtration/HVL

Annual ≥ 2.5 mm Al Al sheets

Quality Control: Exposure Reproducibility

Annual +/- 5% Dosimeter

Quality Control : FS size spatial resolution

Annual +/- 50% Star test pattern Slit or pinhole camera

Display noise

Any patterns or electrical fluctuations that interfere with the detection of the true signal

Image Display (monitors): Luminance (rate of light)

Better than 5% Photometer

DR detectors: Field Uniformity

Daily to Semi-annual Erase plate look for artifacts Expose detector @72" SID (or at 78")

Focal Spot Blur

Degradation of the focal spot can produce blurred structures on the radiograph.

DR detectors: Erasure Thoroughness

Erase plate look for Ghosting or Image Lag For CR - use Al step wedge

Resolution

Is Line pairs per millimeter (Lp/mm). The higher the # of line pairs the better the resolution

Collimation

Light Radiation Field Congruence (Beam Alignment) light field and radiation field must coincide

DR detectors: Spatial Resolution (blurring)

Look for distortions Wire mesh on detector or Standard Bar test

Radiographic quality control

Periodic monitoring of X-ray tube, the associated electric circuits, the accuracy of the exposure factors, and the film processor.

Two types of Image Display Monitors

Primary Secondary (Clinical)

Exposure Reproducibility

Radiation output (measured in mR) should be at the same for a series of exposures in which all technical factors are the same

Quality Control: Collimation

Semiannual or Annual +/- 2% SID Collimator test tool

Image Display (monitors): Luminance Uniformity

Should not deviate more than 30% Photometer (measure 4 corners and center)

SMPTE

Society of Motion Picture and Television Engineers

Image Display (monitors): Display Noise

TG18-AFC (view at 30 cm)

Exposure Linearity

The ability of a unit to produce a constant radiation output from various combinations

Ghosting or Image Lag

When electrical charge is trapped in amorphous silicon or selenium and released over time

Image Display (monitors): Luminance Response (brightness of light)

Within 10% SMPTE test pattern

Stuck pixels appear as _ on a solid white field

black

Luminance uniformity refers to the consistency of a single _ level displayed across a screen.

brightness

Luminance response refers to the various shades of _ from a _.

brightness, test pattern

SMPTE - designed a generic test pattern which can test for most _ qualities

image

Diffuse Reflectance is _ that is randomly _on the digital dislay

light, scattered

Specular Reflectance resorts in _ images of light sources surrounding the display monitor.

mirror

AAMP - designed a test pattern TG18-AFC to measure _ levels

noise

Luminance is the _ of light emitted from a source

rate (intensity)

Photometers

read in units of lux, lumens, or candela/m² (cd/m²)

AAPM- designed a test pattern TG18-CX to check for _on a digital screen

resolution

Dead pixels appear as _ on a solid background

white


संबंधित स्टडी सेट्स

CHapter4 Quiz: Primerica Missed Qz Quest.

View Set

Learnsmart Anatomy and Physiology Chapter 1: Human A&P

View Set

Chapter 11 The Anatomy and Function of the Diencephalon

View Set

Unit 3 - Banking - Financial Management

View Set

Systems Analysis and Design Chapter 1 Questions

View Set

Chapter 10 Change Management Perspectives

View Set

Comprehensive Quiz 14 - Exploring Business Technology

View Set

PrepU Chp 28: Assessment of Hematologic Function and Treatment Modalities

View Set