Chapter 6 X-Ray Tube
Types of external structure supports:
-ceiling support -floor to ceiling support -floor-mount system -Fluoroscopic tube is mounted underneath the x-ray table. -C-arm, a C shaped support for portable fluoroscopy or routine radiography
Small focal spot:
0.5 - 1.0mm in size (300 mA)
Rotating anode provides approximately
1000 times more area for electron beam to interact than a stationary anode
1000 degrees F equals
538 degrees C
Average X-ray tube filament life expectancy is
6-9 hours = 10 to 20 thousand exposures
isotropically
Equal intensity in all directions
filaments will never be used at the same time. T OR F
True
stator and rotor
What are induction motors composed of?
positive side of x-ray tube
anode
most common anode materials for heat dissiaption
graphite, copper, molybdenum
what is the most common cause of tube failure?
vapor induced tube arcing
Functions of the anode
-electrical conduction -mechanical support for target -thermal conduction
Maximum leakage radiation allowed in the x-ray tube is
1 mGy a /h at 1 meter when operated at maximum conditions
Housings have heat capacities in range from
1 to 1.5 million heat units
Large focal spots
1.0 - 2.0mm in size (400 mA and up) if small is used it will be exceed in heat capacity
Heat units for three phase units: three phase six pulse =
1.35 x KvP x Ma x sec.
three phase twelve pulse =
1.41 x KvP x Ma x sec.
the anodes of high-capacity x-ray tubes rotate at
10,000 rpm
Leakage radiation limit
100 milliRoentgens per hour (mR/hr) at 1 meter (m)
Most common anode angle used is ________. Any smaller an angle can limit the primary beam field at shorter SID's.
12 degrees, the small the angle the better it will become
steel: melting point is
1370 degrees celcius
the glass or metal enclosure is relatively large, perhaps
30 to 50 cm long and 20 cm in diameter
Most rotating anodes revolve at ________ .
3400 rpm (revolutions per minute)
tungsten high melting point
3410 degrees C
how many heat units would result from a an exposure of 88 kVp at 300 mA and 2/10 sec using a single phase x-ray generator?
5,280 HU
when electrons have bombard the target __________% of their kinetic energy is converted to heat
99
one hear unit for a single phase unit equals
=1KvP x 1Ma x 1sec. Hu = KvP x Ma x sec.
Rotor
A rotating electromagnet inside the stator.
tube rating chart (3 types) #2
Anode cooling chart: can determine time required for heated anode to cool Thermal capacity of anode is measured in "heat units".
C-arm support system
Ceiling mounted and provide for very flexible x-ray tube positioning
capacitor discharge
Connecting a charged capacitor across a resistor and so enabling the charge to flow from one plate to another. The charge remaining decays exponentially.
line focus principle
Describes the relationship between the actual and effective focal spots in the x-ray tube. A smaller target angle produces a smaller effective focal spot.
Anode heel effect
Due to the geometry of the angled anode target, the radiation intensity is greater on the cathode side.
floor to ceiling support system
Has a single column with rollers at each end, one attached to a ceiling-mounted rail and the other attached to a floor-mounted rail
tube rating chart (3 types) #3
Housing cooling chart: determines time required to cool tube housing.
tube rating chart (3 types) #1 -Most x-ray units have a built in safety feature that will not allow an unsafe exposure to be made.
Radiographic rating chart: provides a guide regarding the most common technical factor combinations that can be used without overloading the tube.
phenomena where thermionic emission is limited because negatively charged electrons make it difficult for subsequent electrons to be emitted by filament.
Space-charge effect, limits our ability to make a large technique
The target area of the anode stuck by electrons is made of:
Tungsten / rhenium alloy
most common cause of tube failure
Tungsten vaporization with deposition on the inside of the glass enclosure
grid biased x-ray tubes
Uses a negative potential difference at the focusing cup to terminate the exposure
in a line focus principle, when the target angle is made smaller, effective focal spot will be smaller.
X-ray tube target angles vary from between 50 - 200
cloud thermionic emission
a cloud of free electrons
Some housings have oil layer between tube and housing, this serves as
an electrical insulator and a thermal cushion. Heat has to be absorbed or it can destroy
grid based x-ray tubes are used in
angiography, capacitor discharge units and pulsed fluoroscopy
the x-ray tubs positive electrode is th
anode
negative consequence of the line focus principle is?
anode heel effect
Leakage radiation
any x-rays, other than the primary beam, that escape the tube housing, results in unnecessary exposure of the patient and technologist
focusing cup
area on the anode that directs the electrons to the focal spot during x-ray production. in order to do that we run a Negatively charged on focusing cup
Fractional focal spots used for magnification techniques in special procedure tubes can be as small
as 0.1mm
Mammography tubes use molybdenum or rhodium targets
because of their low atomic #
the stem of the anode is
between the anode and the rotor
produce two focal spots by using an anode that contains two different angles.
biangular targets
Large focal spot
can withstand heat produced by large exposures. Associated with a large filament. so used when larger body parts are imaged
the x-ray beam has a greater intensity on the _________ side of the x-ray tube
cathode
internal structure
cathode is the negative side of tube, contains filament and focusing cup.
movement of electrons is always from
cathode to anode
xray tube current is controlled by
changing the filament current
Anode construction
copper, graphite molybdenum.
stationary anode x-ray tubes are used in
dental imaging systems, some portable imaging systems, and other special-purpose units in which high current and power are not required
xrays produce alot of heat and heat
destroys xray equipment
tungsten has a thermal conductivity nearly equal to that of copper. Its therefore an efficient metal for
dissipating the heat produced
protective housing protects against
electric shock and serves as mechanical tube support
Anode is driven by an
electromagnetic induction motor
rotating anode allows
electron beam to interact with a much larger target area, so the heat of the anode is not confined in one spot
the focusing cup is negatively charged so that it
electrostatically confines the electron beam to a small area of the anode
x-ray tube is very heavy and requires a sturdy support structure to support its weight.
external structure
X-rays produced outside the target/focal spot. Also referred to as Off Focus
extrafocal radiation
the larger the effective focal spot the greater the spatial resolution. true or false
false it is small the local spot the greater the spatial resolution
usually made of thoriated tungsten
filament
The focusing cup is made of:
filament is embedded in metal focusing cup
The small focal spot will provide a _________
finer detail of image or superior spacial resolution. Associated with small filament
the area of the anode from which x-rays are emitted is known as the
focal spot
refers to an element in the tube that acts as a switch
grid
certain types of x-ray tubes designed to be turned off and on very rapidly
grid-controlled tubes
floor-to-ceiling support system
has a single column with rollers at each end, one attached to ceiling mounted rail and the ohter attached to a floor-mounted rail
Tungsten's high atomic # (74) results in
high efficiency x-ray production.
what is possible with rotating anode?
higher tube currents and shorter exposure times
Virtually all high capacity tubes now constructed use tubes enclosed
in metal. They are moving away from glass and pyrex.
an decrease in the degree of anode angulation equates to a/an __________ in the anode effect
increase
in order to increase the energy of X-ray photons produced at the tube, the kinetic energy of projectile electrons must be increased, this is accomplished by
increasing kVp
the heat capacity can be further improvised by
increasing the speed of anode rotation
The anode spins using an
induction motor
Stator
induction-motor electromagnets that turn the anode
Difference in __________ can vary as much as 45%.
intensity across useful beam
Effectiveness of focusing cup is determined by
it's size and shape, it's charge and filament size and shape, and the position of the filament in the focusing cup
small the angle the
larger the heel effect
The protective housing is made of what?
lead-lined cast steel
The protective housing controls leakage and scatter radiation through the tube window to radiation of
less than 1mGya/h at 1 m
anode heel effect is radiation intensity is greater at cathode side of x-ray beam. This is a by product of
line focus principle
Single excessive exposure can result in
localized surface melting and pitting of anode / target.
Tube failure can occur from
long exposure times
x-rays that are emitted through the heel of the target
lose intensity
the cathode has its own circuit of
low voltage and ?
Space charge effect limits tubes to
maximum Ma ranges from 1000 - 1200
the protective housing also provides __________________ for the x-ray tube and protects the tube from damaged caused by rough handling
mechanical support
As tungsten vaporizes in glass tube arcing can be a problem,
metal tubes maintain a constant electric potential have a longer life and fail less (metal enclosure tubes)
Dual focus tube
most x-ray tubes have two focal spots, small and large (focal track)
Biased focusing cup uses __________ to focus electron stream as it moves toward the anode
negative voltage
Cathode is _____ charged portion of the tube
negatively
anode is _______ charged portion of the tube
positively
What has been material of choice for tube construction?
pyrex
X-ray intensity
quantity of x-rays, how may photons you have
Tube heat must be dissipated through what?
radiation, conduction, convection
ir order to be used more efficiently by the x-ray tube, alternating current is changed to pulsating direct current by the
rectifiers
the stem of the anode is narrow to
reduce thermal conductivity. The stem is usually made of molybdenum bc it is a poor heat conductor
extrafocal radiation is not useful radiation, it
reduces image contrast and extends the size of the focal spot
filaments in a biangle x-ray tube have been placed "end to end" with the
small focus filament above the large filament
the phenomena that makes it difficult for subsequent electrons to be emitted by the filament because of electrostatic repulsion os referred to as
space charge effect
Line-focus principle simultaneously improves
spatial resolution of the x-ray beam as well as the heat capacity of the anode
two types of anodes
stationary and rotating
we produce electrons and we apply energy to them and move them through
the anode and cathode
in rotating anode tubes what is the target
the entire rotating disk
focusing cup must always maintain a greater negative charge than
the filament itself now matter what amperage we use. It must always be greater
in a Grid-controlled x-ray tube
the focusing cup is the grid and therefore the exposure switch
saturation current
the point in which a further increase in KvP will not result in a higher Ma because all the available electrons have been used. (you used your kvp so much that you used all your electrons)
thermionic emission
the release of electrons from the tungsten filament when the electrical current passes through it and heats the filament
Tube window is section of tube designed to let x-rays pass through. These x-rays going through the collimator and onto the body parts are called
the useful beam
the source of electrons within the x-ray tube is the result of which of the following processes
thermionic emission
general purpose x-ray tubes use the rotating anode bc
they must be capable producing high-intensity x-ray beams in a short time
What is the name of the tube failure where vaporized tungsten is collected on the envelope?
tube arcing
the mot common cause of x-ray tube failure is
tube arcing
Glass envelope
tube is evacuated and contains two electrodes (cathode/anode)
tube failure
tube life is extended by using appropriate radiographic factors of Ma, KvP and time.
this relationship between the filament current and the tube currently depends on the
tube voltage
in stationary anode tubes the target consists of
tungsten alloy embedded in the copper anode
most rotating anode x-ray tubes have
two filaments mounted in the cathode assembly "side by side", creating large and small focal spot sizes
xrays emitted through the window are called
useful beam
Target surface irregularities result in
variable and reduced radiation output
preferred detent position
when the x-ray tube is centered above the examination table at the standard SID
Protective housing
x-rays are emitted isotropically, with equal intensity in all directions.
tube failure can be caused by
• Excessive heat for long periods (thermal characteristics) • long exposure times • heating of neck, rotor are & bearings • bearing damage • cracked envelope • excessive MA • vaporized plating • open filament • oil overheating • patient danger