Fresenius water system
action limit for endotoxin
0.125 EU/ml
the hard water limit after the softener is
10 ppm
the delta pressure across a softener should not exceed
10 psi
TDS settings on the RO should be ___ times the normal product TDS
2
action limit for bacteria is
20 cfu/ml
the RO pre filter protects the RO machine from particles up to
5 micron
if RO rejection rate reaches _____ the membranes must be replaced
80%
the over temp alarm is set at
90 deg F
multimedia sediment filter
AKA sand, depth or bed filters, remove large particles, contains multiple layers with each layer retaining smaller particles as small as 10 micron
booster pump
Boosts feed water pressure
deionization tank
DI is emergency backup for RO, rarely used as primary treatment, uses beds of resin beads to remove ions, these are exchanged for hydrogen and hydroxide ions to form water
brine tank
Holds salt pellets mixed with water for the softener
delta pressure
Pressure difference pre and post filter, limit 10psi
a decrease in temperature will decrease the efficiency of an ____
RO membrane
_____ is used to measure total dissolved solids in the water
TDS meter
bicarbonate
a buffer, neutralizes acids that form when the body breaks down proteins
colony forming units
a measure of the number of living bacteria
total chlorine testing algorithm
a tool used by staff in guiding them thru routine testing and what to do if results are at or above 0.10 ppm
endotoxin is
a toxic substance bound to the bacterial cell wall and released when the bacterium ruptures
sepsis is
a whole body inflammation caused by severe infection
conductivity
ability of a fluid to transfer electrical charge
UV light
administers a lethal dose of radiation to eradicate bacteria
RPC ultra low total chlorine strips
another way of testing total chlorine, backup for hach colorimeter
alternative water supply plan
approved by the medical director, provides instructions if city water is interrupted, insures facility meets AAMI standards even if one or more components of the locations system fail, posted in water treatment area
chlorine and chloramines
are harmful to patients, causing hemolysis, together chlorine and chloramines form total chlorine
AAMI
association for the advancement of medical instrumentation, develops voluntary standards for dialysis
frequency of routine total chlorine testing
before the first patient treatment and not to exceed every four hours
SDS consists of
bicarb mix tank, bicarb and acid head tank, acid storage tank, and distribution loops
aluminum and flouride can cause
bone disease
A TDS meter should be ______ weekly
calibrated
chloramines
chlorine + ammonia
carbon tank
contains granular activated carbon which absorbs low molecular weight particles from water, chlorine, chloramines, pesticides, industrial solvents and trace organic substances
sediment filter cartridge
cylindrical cartridge of filter media with a central; drainage core, filter out particles, carbon fines and water softener resin
pheonix meters must be verified ____ with standard solutions and calibrated if they show more than ____ +/- the standard
daily, 0.1
microbial testing
determines if microorganisms (endotoxins and bacteria) are present
filter
devices that remove particles by passing fluid through pores
high chlorination of the water supply testing
done when the city adds extra chlorine in the water supply, due to break in main, flooding or bacterial contamination. GAC testing increased to hourly
total chlorine
free chlorine (hypochlorite ion AKA bleach) + combined chlorine (chloramines) safe level 0.00 to 0.09 ppm
3 characteristics of bacteria and or viruses
gram positive bacteria turn purple when gram stained, can form biofilm slime that protects them from disinfection, microscopic single celled
GAC
granulated activated carbon filters are the primary method to remove total chlorine
SDS may be ____ or _____ fed
gravity or pressure
_____ are used to test water for levels of calcium and magnesium
hardness test strips
product water
has been forced thru an RO membrane
HMIS
hazardous materials identification system, this permanent labeling required on SDS equipment uses colored bars, numbers and symbols to alert staff to chemical hazards
captive air tank, aka expansion or bladder tank
helps maintain minimum system pressure while the booster pump cycles
the ____ should always be set to alert staff on the floor that there is a chance of running out of water
holding tank low level alarm
head tanks
in gravity feed solution delivery systems, concentrates are pumped into this equipment for delivery to patient stations
IFU
instructions for use
water softener works by a process known as
ion exchange
loop pumps are always running to
keep water moving in the loop and minimize bacterial contamination
15 minutes
length of time RO has to be running B4 water samples can be taken for RO water sample and chlorine testing
action level
level at which we must take corrective measures.
TCL-2
log used for breakthrough documentation when total chlorine levels are 0.10 ppm or greater after the first carbon tank
TCL-1
log used for routine daily total chlorine documentation of safe ranges between 0.00 and 0.09 ppm
_____ pumps under the holding tank move water from the tank out to the stations
loop
backwashing
means forcing water backwards thru a filter, removes particles clogging filter
adsorption
means to attract and hold on a surface. the binding of chlorine and chloramine molecules to the granulated activated charcoal
temperature blending valve
mixes hot and cold water to 77 deg F,
routine culture testing should be done monthly prior to ____
monthly disinfection
RO pre filters are changed when
monthly or when delta pressure exceeds 10 psi
the acid feed injection pump uses what kind of acid
muriatic or sulfuric
empty bed contact time (EBCT)
period during which feed water must remain in contact with the charcoal bed in a carbon tank, AAMI standard is 10 minutes total EBCT, fmc 12 minutes
granuflo
powdered acid concentrate, or the liquid acid concentrate after reconstitution
reverse osmosis
pressure is applied to a membrane overcomes the osmotic pressure of the solution, the water then travels across the membrane in reverse of the natural occurrence of osmosis
the over temp alarm is in place to
prevent damage to the RO membranes
_____ diverts the product water to the drain if the water exceeds the TDS alarm limit on the RO
product divert
the water exiting the RO is called
product water
percent rejection
provides an indication of how effective an RO machine is, must be 95% when installed and when membranes are changed, action must be taken when there is a radical change in rate
_____ are used to test for residual chlorine after disinfection
residual chlorine strips
precipitate
solid particles that settle out of a solution and can clog pipes or damage components
SDS
solution delivery system, used to mix and deliver dialysate concentrate solutions to single patient dialysis machines.
the hach meter must be verified daily with a ____
standard spec kit
a TDS meter should be ____ daily
standardized
feed water
tap water before treatment
breakthrough
test results after the first carbon filter that are at or above the action level of 0.10 ppm
low range (LR)
the hach colorimeter should always be set in LR
ion exchange
the process that occurs inside a deionizer for water treatment. unwanted ions are exchanged for hydrogen and hydroxyl ions to create pure water
initial volume
the quantity of treated water in the mix tank prior to adding bicarbonate concentrate powder
total volume
the quantity of treated water plus the NaturaLyte bicarbonate base powder required to produce bicarbonate base concentrate as stated on the label
fill volume
the quantity of treated water plus the number of bags of acid base powder required to produce the total volume of acid base concentrate as stated by the manufacturer
microns
the unit of measurement for filter pores.
chloramine adsorption is affected by ____, ___ and ____
time, pH, temperature
backflow prevention device
to keep water from flowing backwards into incoming water supply, installed at start of system
0.00 to 0.09 ppm
total chlorine levels that are patient safe
_____ are located in the distribution loop after the holding tank to remove bacteria and endotoxins
ultra filters
what generally follows the UV light in the system
ultrafilter to catch killed bacteria
ultrafilter
used to control and remove particulates, bacteria, endotoxins, and other microorganisms
temporary labeling
used to label mix tanks with contents that change, must be conspicuous/noticeable
chemical feed system
used to lower feed water pH in RO systems, acid injection corrects alkalinity or pH >8.0.
hydrometer
used to measure specific gravity of an acid concentrate after mixing, must be +/- 2% of expected range
bicarbonate mixer
used to mix bicarbonate concentrate solutions for jugs
water softener
used to reduce the concentration of calcium and magnesium that form scale
hach pocket colorimeter
used to test for total chlorine
spec check color standard kit
used to validate hach colorimeter daily
phoenix meter
used to verify dialysate conductivity
reject water
waste stream sent to drain along with the solutes removed during dialysis
direct feed
water system design in a facility that does not have a holding tank
indirect feed
water system design in a facility that has a holding tank
membrane filters
water treatment cartridges containing thin membranes with pores of a specified size. remove small particles and some solutes
valve panel
where bicarb concentrate, acid concentrate and RO water are centrally delivered for use by HD machines