DIGITAL EXAM 1 122
CR technology uses
Reusable imaging plates, monoenergetic laser stimulation, reader assemblies
When setting kVp levels using DR technologies, it is important to understand that
ALL OF THE ABOVE higher kVp values produce a more narrow histogram kVp levels should not exceed 80 kVp for non grid radiography when scatter ratios reach 50 percent or greater, radiographic grids should be used
The luminescence of blue-violet light from a CR fluorohalide, due to laser stimulation
Is proportional to the x-ray energy exposure
7 Imaging plate layers
▪ Top protective layer ▪ Phosphor layer ▪ Conductor layer ▪ Support layer ▪ Light shielding layer ▪ Bottom protective backing layer
Objectionable quantum noise on an image that makes its acceptance marginal, generally indicates an underexposure of
25 percent
With a variable-kVp exposure technique system, the threshold kVp is typically set at
30 kVp
An x-ray exposure taken at 80 kVp has an average energy of
35 keV
In CR the latent image consists of millions of
F-centers and electron holes
A CR imaging plate that has been grossly overexposed can demonstrate __________ if the erasure process is incomplete
Ghosting
Pixel bit depth determines
Gray shade levels
As you complete your daily, late afternoon duties as a radiographer, you notice a CR cassette that was exposed earlier in the morning. When you processes the cassette in the reader, the resultant image will demonstrate
High levels of quantum noise
In CR technology, preprocessing of data includes
Histogram analysis, grayscale analysis, exposure data recognition
The fluorohalides used in CR are predominantly composed of
Bromide
Select the TRUE statement regarding computer radiography
CR imaging plates are very sensitive to post-exposure radiation
The area of a DR detector DEL that is insensitive to x-ray exposure.
Consists of DEL electronics.
To reproduce an analog signal with truer representation and accuracy, the most effective method to do so is
Increase the sampling rate to a value greater than the Nyquist value
All of the following are true of Moire effect EXCEPT
It is most likely to occur with DR technology, particularly amorphous selenium
Underexposure using DR detectors will result in A. less spatial resolution. B. photon saturation. C. quantum noise. D. electronic noise.
Quantum noise
Readout of CR data from the imaging plate is a line-by-line process which creates a
Raster pattern
As a radiographer you are looking at an image you had just taken on the review monitor the image looks marginally noisy youre next step should be to
Refer to the exposure index # to confirm image appearance
As a radiographer, you are using Dr systems that displayed EI# as part of the DICOM header the EI# are calculated as directly proportional and linear to the exposure
Repeats the exposure at 85 kvp @ 20 mas
In CR, pixel pitch is dependent upon
Sampling frequency
in a digital detector the term fill-factor refers to the
Size of the DEL
The tendency of technologists to use more exposure than needed, and allowing computer rescaling to compensate for it, has lead to
dose creep
In the technology of CR, the latent image is actually
electron "holes" in Europium
When adding annotation to a digital radiographic image, it is important to remember that
electronic annotation for patient laterality is unacceptable
Elimination of extraneous light from the edges of a digital radiographic image, is achieved through
electronic masking
a key feature of DR/CR technologies is the increased sensitivity of the receptor to a wider range of remnant radiation energies, when compared to conventional film- screen systems. This is referred to as
low contrast resolution
Computerized digital images consist of a ________ of picture elements, each with a video display value or gray shade.
matrix
The component that converts light to an electronic signal in a digital detector is the
photoconductor
Computed radiography uses (a) _____ for latent image formation.
photostimulable phosphor
With digital technologies (DR/CR), perhaps the single-most important consideration when using this technology is
the total amount of exposure incident upon the detector
a major distinsction b/n DR and CR receptors is that
*CR requires particular attention to placement of the anatomy on the receptor *DR is not affected by collimation symmetry *The latent image in CR is time-sensitive
Digital radiography first began to be used in clinical settings in the _____.
1980's
Fixed kvp techniques systems require
An optimum threshold kvp
In computer radiography, is the reader scanning frequency and the radiographic grid frequency are similar and in the same orientation, what is likely to occur?
Moire effect
The response to x-ray exposure with CR imaging plates is
Nonlinear
Computed radiography is also referred to as
PSP technology
Underexposed radiographic image demonstrates
Photon starvation
Positioning of an AP wrist protection off center using a CR receptor, will likely lead to
Preprocessing errors of the data, an incorrect histogram analysis, sub optimum image quality
When using CR technology, it is important to remember that
The IP is extra sensitive to exposure after the initial x-ray exposure, the IP will lose about 25% of its signal in 8 hours if not processed, the IP is very sensitive to scatter, making grids very important ALL OF THE ABOVE
With digital detector technology, the traditional rules of exposure and image quality do not necessarily apply completely. One of these rules is that
The amount of scatter formation is primarily a function of mAs rather than kVp
What is the relationship between CR plate size and spatial resolution?
The best CR spatial resolution is achieved with the smallest size CR plates
With digital detector technology, the electronic signal is:
The latent image An analog signal A data set of signal intensities (all of the above)
When using a Fuji CR system in the fixed mode of operation
The system will not correct for exposure inadequacies and adjust for density problems, your exposure techniques must be accurate, similar to film/screen technology BOTH B AND C
DR panels that are connected to the radiographic unit with a "hard wired" connection cable
are considered to be "tethered"
Select the TRUE statement with regard to complementary metal oxide semiconductor (CMOS) technology:
They are not widely used with flat panel DR technology
A common feature between indirect and direct DR technology is
Thin film technology (TFT)
In direct digital detectors, the electronic readout of signal is accomplished by the
Thin film transistor (TFT)
True/False Csl/CCD is a indirect DR process by which x-rays are converted first to light and then to electric signal.
True
All of the following are related EXCEPT A. Automatic B. Variable C. Semi- Automatic D. Fixed
Variable
The purpose of a charge couple device (CCD) when used with a DR detector is to
act as a photoconductor collect the light from the scintillator and convert it to the electrons (Both B and C)
Extraneous information that appears on a medical image, that does not represent the patient and their medical condition, is referred to as
an artifacte
Of the choices below, select the DR panel that would provide the highest spatial resolution. a. DEL size 200 microns, 75% fill factor b. DEL size 125 microns, 95% fill factor c. DEL size 110 microns, 75% fill factor d. DEL size 140 microns, 95% fill factor
b. DEL size 125 microns, 95% fill factor
A radiographic image that demonstrates image receptor electronics on the final image superimposed over patient anatomy is likely the result of
backscatter radiation
DR panels have published specifications from the manufacturer that outline performance conditions. DR panels are particularly vulnerable to
bodily fluids
Indirect DR uses ________ as the photoconductor material. a. calcium tugstate b. amorphous selenium c. amorphous silicon d. gadolinium oxysulfide
c. amorphous silicon
Indirect DR detectors employ the use of _____________ as a scintillator. a. cmos b. zinc cadmium sulfide c. cesium iodide d. ccd
c. cesium iodide
Conventional x-ray tubes and cassettes are used with _____ radiography systems
computed
DR detector element saturation can result in image
data drop
Extreme overexposure to a DR detector can result in
detector DEL saturation
the calculation of exposure index values can vary between manufacturers of DR/CR technology. the AAPM has established a scale of exposure ranges known as a(n) ___, that accounts for these variations in calculations
deviation index (DI)
The total amount of clinically significant diagnostic image on a medical image is referred to as
diagnostic yield
The relationship between the detector "fill factor" and spatial resolution is:
direct
A DR detector that is integral to the x-ray system means that the detector
is built in to the design of the x-ray equipment and the radiographer cannot access it.
Auto-detection technology with a DR panel
is often referred to as a trigger panel
CR systems acquire clinical information by
laser scanning a PSP plate
The final appearance of a digital image in terms of brightness and contrast is principally due to
the application of a look up table
Which of the following factors would decrease the remnant radiation exposure to a dr detector
using a 8" air-gap.
Essential to ALARA compliance is
using accurate exposure techniques based upon patient measurement
DR and CR image receptors
work well with automatic exposure control.