EXAM 3-Principles of Radiographic Imaging
If a shoulder examination is done using an 8:1 ratio grid and 12 mAs, how much mAs should be used if the examination must be done tabletop (no grid)? a. 1.5 mAs b. 3 mAs c. 48 mAs d. 96 mAs
3 mAs
The automatic collimator: a. automatically collimates to the anatomy of interest b. is required by law on all new equipment c. is seldom found on modern x-ray equipment d. automatically collimates to the IR size
automatically collimates to the IR size
The distance between the grid and the convergent line is the: a. convergent line b. focal distance c. convergent point d. focal range
focal distance
The relationship between the height of the lead strips and the distance between them is the grid: a. number b. ratio c. frequency d. Rating
ratio
If 24 mAs and a 6:1 grid results in an excellent radiograph, how much mAs is needed with a 5:1 grid? a. 16 mAs b. 20 mAs c. 28.8 mAs d. 36 mAs
16 mAs
The purpose of automatic collimation is: a. improved image quality b. limiting patient exposure c. increasing the cost of equipment d. none of the above
limiting patient exposure
To compensate for the reduction of radiation reaching the IR when a grid is used, _______ is typically increased. a. SID b. kVp c. mAs d. OID
mAs
Beam-restricting devices are located just above the tube housing. a. true b. false
false
Grids absorb almost all of the scatter radiation and allow all of the transmitted photons to pass through to the IR. a. true b. false
false
It is safe to assume that the x-ray beam field size and the light field size are exactly the same. a. true b. false
false
The air gap technique is a frequently used method for reducing the amount of scatter radiation reaching the IR. a. true b. false
false
The grid limits the amount of scatter radiation produced in the patient. a. true b. false
false
The grid used for the air gap technique should be high frequency. a. true b. false
false
When using automatic collimation, collimating to the anatomy of interest is an unnecessary waste of time; the beam is already collimated. a. true b. false
false
Absorption of transmitted photons by a grid caused by misalignment is: a. grid focus b. grid reduction c. grid cutoff d. transmission reduction
grid cutoff
Anyone in the room during an exposure must wear a lead apron to absorb the scatter radiation produced in the: a. table b. patient c. air d. beam
patient
For lateral lumbar spine, placing a lead shield on the table behind the area of exposure: a. reduces the scatter exiting the patient b. reduces the scatter reaching the IR c. A and B d. none of the above
reduces the scatter reaching the IR
The unrestricted primary x-ray beam is: a. round b. oval c. square d. rectangular
round
Adding a grid or increasing grid ratio results in an increase in patient dose. a. true b. false
true
Focused grids are more effective than parallel grids in absorbing scatter radiation and allowing transmitted radiation to reach the IR. a. true b. false
true
Scatter radiation is of no value in radiographic imaging. a. true b. false
true
If a table top exposure (no grid) of a humerus requires 2.5 mAs, how much mAs is needed if a 12:1 ratio grid is added? a. 0.21 mAs b. 0.5 mAs c. 12.5 mAs d. 30 mAs
12.5 mAs
If 22 mAs and a 5:1 grid results in an excellent radiograph, how much mAs is need with a 12:1 grid? a. 8.8 mAs b. 9.2 mAs c. 52.8 mAs d. 55 mAs
55 mAs
During the exposure, the grid in the bucky: a. reciprocates b. moves slightly from side to side c. moves slightly from top to bottom d. A and B
A and B (reciprocates & moves slightly from side to side)
Restriction of the primary beam to the anatomy of the interest: a. reduces patient exposure b. is a major cause of repeated images c. reduces the amount of scatter produced d. A and C
A and C
A grid with lead strips that run parallel to each other is a: a. a parallel grid b. focused grid c. nonfocused grid d. A and C
A and C (a parallel grid & nonfocused grid)
The Moire' effect can be caused by: a. using a grid with a frequency similar to the CR laser scanning frequency b. using a focused grid upside down in the bucky c. using a grid cassette in the bucky d. A and C
A and C (using a grid with a frequency similar to the CR laser scanning frequency & using a grid cassette in the bucky)
Grids are typically only used when the part measures more than ________ cm and __________ kVp is need for the examination. a. 5; 50 b. 10; 50 c. 10;60 d. 15; 65
10; 60
A linear grid: a. has lead strips that cross each other b. has lead strips that travel in one direction c. is the most common type of radiographic grid d. B and C
B and C (has lead strips that travel in one direction & is the most common type of radiographic grid)
For radiographic procedures, scatter radiation is primarily the result of: a. photoelectric interaction b. pair production c. Compton interaction d. classical scattering
Compton interaction
The formula for the grid conversation factor is: a. GCF= mAs without grid/ mAs with grid b. GCF= mAs with grid-mAs without grid c. GCF = mAs with grid/ mAs without grid d. GCF= mAs without grid-mAs with grid
GCF = mAs with grid/ mAs without grid
Which of the following is the formula used to determine grid ratio? a. Grid ratio=distance between the lead strips/ height of the lead strips b. Grid ratio= width of the lead strips/ height of the lead strips c. Grid ratio= height of the lead strips/ distance between the lead strips d. Grid ratio= height of the lead strips/ width of the lead strips
Grid ratio= height of the lead strips/ distance between the lead strips
The air gap technique is based on creating a gap by increasing the ___________. a. SID b. SOD c. OID d. all of the above
OID
The exam room in our lab that offer automatic collimation is: a. Room 1 b. Room 2 c. Room 3 d. Room 4
Room 3
To compensate for some of the loss of recorded detail due to the large OID used with the air gap technique, the __________ should be increased as well. a. grid ratio b. SID c. focal spot size d. A and B
SID
Grid cutoff results in: a. reduced exposure to the IR b. a light film-screen image c. possible quantum mottle d. all of the above
all of the above
The volume of tissue irradiated is determined by: a. the area of collimation b. patient thickness c. field size d. all of the above
all of the above
Which of the following factors needed to be considered when determine the grid to be used for a specific examination? a. patient exposure b. level of contrast improvement needed c. focal range d. all of the above
all of the above
When imaging the pelvis, changing from 14 x 17 collimation to 8 x 10 collimation requires: a. an increase in kVp b. a decrease in kVp c. an increase in mAs d. a decrease in mAs
an increase in mAs
The beam-restriction device that results in images with the most unsharpness around the edge of the image is the: a. aperture diaphragm b. collimator c. cone or cylinder d. none of the above
aperture diaphragm
The flat piece of lead with a hole in it is the: a. aperture diaphragm b. collimator c. cone or cylinder d. none of the above
aperture diaphragm
The higher the grid ratio, the ____________ the scatter clean up and the ____________ the radiographic contrast. a. better; higher b. worse; lower c. better; lower d. worse; higher
better; higher
The grid is located: a. just below the x-ray tube window b. between the patient and IR c. just below the IR d. none of the above
between the patient and IR
The reason the grid in the bucky reciprocates is to: a. absorb more scatter radiation b. allow more transmitted radiation to reach the IR c. blur out the grid lines d. all of the above
blur out the grid lines
The grid is part of a device located just below the tabletop that also includes a tray to hold the IR. The device is the: a. wafer grid b. bucky c. grid cap d. grid cassette
bucky
The beam-restriction device that includes a visible light projecting the x-ray field size onto the patient is the: a. aperture diaphragm b. collimator c. cone or cylinder d. none of the above
collimator
The most effective and often used radiographic beam-restriction device is the: a. aperture diaphragm b. collimator c. cone or cylinder d. none of the above
collimator
A flat piece of lead with a hole in it that has a flange directed toward the patient is a(n): a. aperture diaphragm b. collimator c. cone or cylinder d. none of the above
cone or cylinder
If all the convergent points were connected along the length of the grid they would from the: a. convergent line b. focal distance c. convergent point d. focal range
convergent line
The location in space where all the lines extended from the lead strips meet is the: a. convergent line b. focal distance c. convergent point d. focal range
convergent point
Beam restriction serves which two main purposes: a. decrease patient dose and decrease scatter b. decrease patient dose and increase scatter c. increase patient dose and decrease scatter d. increase patient dose and increase scatter
decrease patient dose and decrease scatter
____________ collimation means __________ field size. a. decreasing; increasing b. increasing; increasing c. decreasing; decreasing d. none of the above
decreasing; increasing
The recommended SIDs that can be used with a focused grid is the: a. convergent line b. focal distance c. convergent point d. focal range
focal range
The number of lead lines per inch or centimeter is the grid: a. number b. ratio c. frequency d. rating
frequency
The type of grid that includes a channel for the IR to slide into is the: a. wafer grid b. bucky c. grid cap d. grid cassette
grid cap
An IR that has a grid permanently attached to its front surface is a: a. wafer grid b. bucky c. grid cap d. grid cassette
grid cassette
To determine the amount of mAs to be increased when making grid changes, it is necessary to use the GCF, also known as the: a. grid constancy figure b. grid compensatory factor c. general compensation factuals d. grid conversion factor
grid conversion factor
A short-dimension grid: a. is typically used for most examinations b. has lead strips running perpendicular to the long axis c. has lead strips running parallel to the long axis of the grid d. A and B
has lead strips running perpendicular to the long axis of the grid
Which of the following produces the greatest amount of scatter radiation exiting the patient? a. low kVp and small field size b. low kVp and large field size c. high kVp and small field size d. high kVp and large field size
high kVp and large field size
The formula to use when changing from one grid to another is: a. mAs1/mAs2= GCF2/GCF1 b. mAs1/mAs2= (GCF2)^2/GCF1)^2 c. mAs1/mAs2= GCF1/GCF2 d. mAs1/mAs2= (GCF1)^2/GCF2)^2
mAs1/mAs2= GCF1/GCF2
The air gap technique reduces the amount of scatter radiation reaching the IR because: a. more scatter is absorbed in the patient b. less scatter is produced in the patient c. more scatter misses the IR d. less scatter misses the IR
more scatter misses the IR
When the x-ray beam is not lined up with the center of the grid you have ________ cutoff. a. upside-down focused b. off-level c. off-center d. off-focus
off-center
When the SID is above or below the focal range you have _________ cutoff. a. upside-down focused b. off-level c. off-center d. off-focus
off-focus
The only grid cut-off error that can occur with a parallel grid is: a. upside-down focused b. off-level c. off-center d. off-focus
off-level
When the x-ray beam is angled across the lead strips you have ________ cutoff. a. upside-down focused b. off-level c. off-center d. off-focus
off-level
The material between the lead strips of a grid must be: a. radiopaque b. radiolucent c. made of lead d. made of tungsten
radiolucent
When there is significant collimation, the amount of radiation reaching the IR is: a. reduced b. increased c. the same d. none of the above
reduced
A focused grid's lead strips are angled to match: a. the anode angle b. the angle of the x-rays in the primary beam c. the angle of the scattered photons leaving the patient d. the tube angulation
the angle of the x-rays in the primary beam
The probability of a Compton interaction occurring is related to: a. the energy of the x-ray photon b. the atomic number of the matter c. the mAs used d. all of the above
the energy of the x-ray photon
Linear grids are the most commonly used because: a. they are very inexpensive b. the tube can be angled in all direction c. the tube can be angled in the direction of the lead lines d. all of the above
the tube can be angled in the direction of the lead lines
Grids absorb scatter radiation because the scattered photons: a. have high energy b. have extremely low energy c. travel in the same direction as the transmitted photons d. travel at an angle to the transmitted photons
travel at an angle to the transmitted photons
The cutoff that appears as a strip of appropriate density in the middle of the image and a significant loss of density on both sides is: a. upside-down focused b. off-level c. off-center d. off-focus
upside-down focused
The two major factors that affect the amount of scatter radiation produced and exiting the patient are: a. kVp and mAs b. volume of tissue irradiated and mAs c. volume of tissue irradiated and kVp d. kVp and the use of a grid
volume of tissue irradiated and kVp
The type of grid that is usually taped to the front of an IR is the: a. wafer grid b. bucky c. grid cap d. grid cassette
wafer grid
The Moire' effect is also known as the _________ pattern. a. cheetah b. giraffe c. lion d. zebra
zebra