Chapter 5

Ace your homework & exams now with Quizwiz!

Percentage of three phase, 6 pulse

13%

Most X-ray systems operate on:

220 V power

440 volts is supplied to 1000 primary turns of an auto transformer. If the desired output voltage is 100 V how many secondary turns must be tapped?

227

Filaments operate at a current of _____ A

3-6 A

A filament current of 5 A is necessary for thermionic emission. What electron flow is this?

3.2 x 10 to the 19

Percentage of 3 phase, 12 pulse:

4%

220 V is supplied to 800 primary turns of an autotransformer. What will be the output voltage across 200 secondary turns?

55 V

To generate three phase 6 pulse power at least how many rectifiers are necessary:

6

Percentage of full wave rectifued

71%

The main advantage of full wave rectification over half wave rectification is:

A greater number of x-rays per cycle

In the design of an auto transformer:

A single coil serves as both the primary and secondary coils

Unrectified voltage

Across an x-ray tube would cause it to act like a diode, conducting electricity in only one direction

The principal purpose of the auto transformer is to:

Adjust voltage

Transformers only operate on _____ current

Alternating (AC)

Equation Watts=

Amperes X volts

Xray tables are most made of:

Carbon fiber

A rectifier:

Changes AC to DC

The auto transformer has only one:

Coil

Space charge:

Collection of electrons around filament

Which of the following would be correct to use for expressing X-ray tube current?

Coulombs/second

Equation Power=

Current X potential

Power rating=

Current X potential

The auto transformer can be used to do which of the following?

Increase kVp

Which of the following is an advantage of three phase power over single phase power?

Increased x-ray intensity per mAs

Semi conductor rectifier:

Is a solid state device

N-type:

Loosely bound electrons are free to move

kVp meter:

Measures voltage output from the autotransformer.

In half with rectification each inverse half cycle in the primary circuit corresponds to how many voltage pulses across the x-ray tube?

One

Autotransformer:

Power supply to the x-ray imaging system

Which of the following is used to determine the voltage before exposure?

Pretending voltmeter

Electeomagnetic or electric motive force:

Production of electrical conductor in a changing magnetic field

The output voltage from the auto transformer is;

Proportional to the turns ratio

We want the table to be:

Radiolucent

Secondary side is a:

Step up transformer

Radiation quantity:

The amount of energy transferred

primary voltage / secondary voltage

Voltage it receives/ voltage it provides

Which of the following is higher for a single phase high voltage generator that for a three phase high-voltage generator?

Voltage ripple

Line compensation:

is required to stabilize voltage

The disadvantage of a self rectified circuit is:

its limited exposure time

Which of the following is directly connected to the auto transformer?

kVp meter

Power rating is expressed in what unit.

kilowatt

X-ray tube current crossing from cathode to anode is measured in:

mA

line compensator:

measures the voltage provided to the x-ray imaging system and adjusts that voltage to precisely 220 V

X-ray tube current is usually measured in which of the following?

milliamperes (mA)

the control of exposure time is always:

on the primary side of the high voltage transformer

Full wave rectification:

requires at least four rectifiers

Power to the primary side of the high voltage transformer comes from the:

secondary side of the autotransformer

The filament transformer is usually:

step-down transformer

high voltage transformer:

step-up transformer

Full wave rectification

the conversion of the opposing half of the incoming electron flow to always move in the same direction, instead of discarding half the cycle

Concerning the transformers used in x-ray circuit:

the high voltage transformer is a step up device

The filament transformer is a ________ transformer.

to step down

Selection of kVp:

uses the step up transformer

Percentage of high frequency:

1%

Percentage of self rectified:

100%

The voltage ripple associated with the various x-ray generators is:

100% for self rectification

A change in voltage waveform from the primary side to the secondary side of the high-voltage transformer produces a change in;

amplitude

Only difference between the primary and secondary waveforms is their:

amplitude

Diode:

an electronic device that contains two electrodes

The filament circuit:

begins at the autotransformer and ends at the filament

The unit mAs:

could be expressed in coulombs

step-down transformer

decreases voltage

Radiation quality:

determines degree of penetration and the amount of energy transferred

Which of the following is contained in a typical high-voltage generator?

diode rectifiers

Oil is used in the high-voltage section of an x-ray imaging system for which of the following functions:

electrical insulation

P-type:

has a "hole" or is missing electron

The voltage supplied to an x-ray imaging system is 220 V the voltage used by the x-ray tube is produced by which of the following?

high voltage transformer

Which of the following is a disadvantage of three phase power compared with single phase power?

higher capital cost

The principal purpose of the high-voltage transformer is to do which of the following?

increase voltage

The filament transformer:

increases current

step-up transformer:

increases voltage

Rectifier:

Device which allows current to flow in only one direction

The auto transformer converts:

Electric energy to electric energy

Falling load generator:

Exposure begins at maximum mA and the mA drops as the anode heats

The autotransformer operates on the principle of:

Faraday's Law

Temperature is controlled by the:

Filament

Exposure technique of 100 mA at 100 ms compared with 50 mA at 50 ms results in:

Four times the total number of electrons

Which of the following principles of boulder to rectification produces the maximum efficiency of x-ray production:

High frequency generator

space charge effect:

a correction circuit has to be incorporated to counteract the

Half-wave rectification

a method of converting alternating to direct current in which half of the current is lost

Operating console:

allows the technologist to control the x-ray tube current and voltage


Related study sets

The 14 Longest words in the english language.

View Set

Ch. 2: Achieving Strategic Fit in a Supply Chain

View Set

Your Rights: Personal Privacy 100% quiz

View Set

Історія України. ЗНО. 6-а тема

View Set

Module 7: Week 7 Interactive Study Guide

View Set

Flashcard Reminder. What is true about Jamestown Settlement

View Set

Chapter 9: Human Resource Management

View Set

BUS LAW CH. 10 Review (CONSIDERATION)

View Set

Acct 2110 cornette test 2 book notes

View Set

Conflict Theory Video Presentation

View Set