Puritan Bennett 840
What two sources of gas powers the PB840
Air, Oxygen 35-100 psig
Rise time %
Available 1-100%: in PC, PS, VC+, VS, BiLevel Purpose to adjust the rate (speed) in which flow reaches from baseline to peak, used to make sure inspiratory needs are met without overshooting PT, slow time may increase WOB (840 defaults to 50%)
Auto-Peep
Expiratory pause key: used to estimate end-expiratory pressure, automatic pause the key is pressed and the ventilator waits until a pressure in circuit is stabilized Manual pause: key is pressed and held cannot be held for longer than 30 seconds.
3 test for PB840
POST: done by vent automatically when vent is on EST: done by service tech SST: takes 3 minutes verifies proper vent operations, compliance, leaks, resistance
Expiratory Sensitivity
Set the percentage of peak flow which the ventilator will cycle out of inspiration, available in spontaneous mode, adjustable from 1-80%, a lower E sens will increase I time, a higher E sens will decrease I time, the proper setting will stop insp flow and allow for exhalation, defaults to 25%
4 function keys
Vent settings: activate sandbox apnea settings: choose apnea setting alarm set up: choose alarms other screen: used to communicate to outside devices
what is the BPS
back up power source; when electrical power drops below a minimum level, BPS will power the vent; Indicator is in the upper right portion of front panel
GUI
graphic user inter phase, 3 functions touch screens, status indicator, knob and selector
what do the indicator colors mean
green light means vent has two minutes of power charged, and is operating, yellow light means BPS is operating and AC power is inaddiquette
COPD and E sens
higher e sens will create longer Te
Sand box
make a lot of mode changes and multiple setting changes, breath timing can be found here, I:E ratio, Ti, Te
Inspiratory pause key
pressing and holding will pause inspiration for 7 seconds, will pause pressure time curve and measure plateau pressure and compliance/resistance
SST
should be done every new PT, every 15 days, new humidifier and circuit, PT cant be connected/"Y"
Patient Data
very top of GUI, Total breaths, I:E, PEEP, Vt
How is auto peep fixed
long expiration, decreasing Vt/RR, lower RAW, bronchodilators