RAD 214 Chapter 4
Higher grid ratio =
higher contrast
Automatic collimators are also called
PBL-positive beam limitation
Recommended SIDs that can be used with a focused grid is the
focal range
Focused grids lead strips are angled to match
the angle of the X-rays in the primary beam
The overall quality of the radiographic image is determined by
photographic properties visibility of recorded detail geometric properties
Spatial resolution is improved with
increased pixel density and decreased pixel pitch
As kVp increases, amount of scatter hitting the IR
increases
Decreasing collimation =
increasing field size
Medium frequency grid
120-150 lines/inch used with stationary grid holders
High frequency grid
150-170+ lines/inch used with stationary grid holders avoids aliasing
Grid that provides excellent scatter clean up but requires close to perfect alignment is
16:1
Long dimension grid
lead strips run parallel to long axis portable chest length wise
Short dimension grids
lead strips run parallel to short axis ei. portable chest crosswise
Purpose of automatic collimator is
limiting patient exposure
Higher grid frequency =
lower contrast
To compensate for the reduction of radiation reaching the IR when a gird is used, ___ is typically increased
mAs
Grid used for air gap technique should be high frequency
false
Grids absorb almost all of the scatter radiation and allow all of the transmitted photons to pass through to the IR
false
The grid limits the amount of scatter produced in the patient
false
With automatic collimation, collimating to anatomy of interest is unnecessary because it is already collimated
false
Xray beam field size and light field size are exactly the same
false
Absorption of transmitted photons by a grid caused by misalignment is
grid cutoff
mAs/relative speed formula
mAs old/mAs new = RS new/RS old
Grid with lead strips that run parallel to each other is a
parallel nonfocused
Volume of tissue is determined by
part thickness and field size (collimation)
When should you use a grid?
part thickness is 10 cm or more kVps over 60/70
Anyone in the room during an exam must wear a lead apron to absorb the scatter radiation produced in the
patient
As collimation increases,
patient dose decreases scatter radiation decreases radiographic contrast increases quantum noise increases radiation reaching IR decreases
As collimation decreases,
patient dose increases scatter radiation increase radiographic contrast decreases quantum noise decreases
What is used to determine the grid used for an exam
patient exposure level of contrast improvement needed focal range
Restriction of primary beam to anatomy of interest
reduces patient exposure reduces amount of scatter
Lateral L-spine, placing a lead shield on the table behind the area of exposure
reduces scatter reaching the IR
The unrestricted primary X-ray beam is
round
Focused grids are more effective than parallel grids in absorbing scater and allowing transmitted radiation to reach the IR
true
Scatter is of no value in xray
true
To overcome the decrease in sharpness caused by air gap technique, increase in SID is required
true
Collimators have
two or three pairs of shutters two way mirror light source face plate with crosshair
Cutoff that appears as a strip of appropriate density in the middle of the image and a significant loss of density on both sides is
upside-down focused
Grids are typically used when the part measures more than ___ cm and ___kVp is needed
10, 60
Low frequency grid
100-120 lines/inch used with moving grid assemblies (bucky)
What type of interaction would occur between a 50kev X-ray photon and an atom with inner shell electrons that have a binding energy of 60 kev or more?
Classical
Scatter is primarily from which interaction?
Compton, 90%
Beam restriction devices are located above tube housing
False
Who invented the grid?
Gustav Bucky in 1913
Who invented the moving grid?
Hollis Potter in 1920
Is it possible for a photoelectric interaction to occur between a 62 kev xray photon and an atom with inner-shell electrons that have a binding energy of 60 kev or more?
No
Air gap technique is based on creating gap by increasing
OID
Image intensification steps
Remnant radiation reaches IR Latent image is processed by the computer Manifest image is displayed on a monitor Manifest image can be adjusted because it is composed of digital data
To compensate for some of the loss of recorded detail due to large OID with air gap, ___ should be increased
SID
Is attenuation the reduction in energy or amount of radiation in the primary beam?
Yes
Function of a grid
absorbs scatter radiation
Grid lead content
affected by grid frequency and ratio
The reason the grid in the bucky reciprocates is to
blur out the grid lines
Reciprocity law
any combo of mA and time that has the same mAs will produce the same quantity of radiation exposure
Beam restriction device that results in images with the most unsharpness around the edge of the image is
aperture diaphragm
Simplest type of beam restricting device is a flat piece of lead with a hold in it is called
aperture diaphragm
Volume of tissue irradiated is determined by
area of collimation patient thickness field size
Automatic collimator
automatically collimates to the size of the IR
What does PBL do?
automatically collimates to the size of the IR and for changes in SID
Higher the grid ratio, ___ the scatter clean up and the ___ the contrast
better, higher
Grid is located
between patient and IR
The grid is part of a device located just below the table top that also includes a tray to hold the IR. This device is the
bucky
Beam restriction device that includes a visible light projecting the xray field size on the patient is
collimator
Most easily adjusted beam restriction device is
collimator
Most effective and often used beam restriction device is
collimator
Flat piece of lead with a hole in it that has a flange directed toward the patient is a
cone or cylinder
If all the convergent points were connected along the length of the gird they would form the
convergent line
Location is space where all the lines extended from the lead strips meet is the
convergent point
Grid cutoff
decrease in the number of transmitted photons reaching the IR because of some misalignment of the grid
Increasing collimation results in
decreased scatter and higher contrast
Diseases that decrease absorption characteristics
destructive
Tube angulation results in
elongation and reduced IR exposure
Distance between grid and convergent line is the
focal distance
Number of lead lines per inch or cm is the grid
frequency
Type of grid that is permanently mounted but includes a channel for the IR to slide into is the
grid cap
An IR that has a grid permanently attached to its front surface is a
grid cassette
To determine the amount of mAs to be increased when making grid changes, it is necessary to use the GCF, aka
grid conversion factor
A short-dimension grid
has lead strips running perpendicular to the long axis of the grid
A linear grid
has lead strips that travel in one direction, most common type of grid
Formula for grid ratio
height of lead strips/distance between lead strips
Grid ratio
height of the grid strip divided by the thickness of the interspace material
What produces the greatest amount of scatter exiting the patient?
high kVp and large field size
As compared with lower kVp, using higher kVp will result in scatter photon that is of ___ energy
higher
The probability of a photoelectric interaction is ___ in bone than in soft tissue
higher
Xray of parts having different tissues types has ___ contrast than that of a part with similar tissue types
higher
Scatter reduces ___ and increases ___
image contrast, dose
Purpose of a grid
increase image contrast
When imaging the pelvis, changing from 14x17 to 8x10 requires
increase in mAs
Increased grid ratio =
increased contrast increased patient dose likelihood of grid cutoff increases
In digital imaging, spatial resolution is improved with
increased pixel density and decreased pixel pitch
Larger field size produces___ scatter
more
Thicker tissue produces___ scatter
more
With a higher kVp, you get
more even penetration of all tissues more scatter hitting the IR (both reduce image contrast)
Air gap technique reduces the amount of scatter radiation reaching the IR because
more scatter misses the IR
Grid Conversion formula
new mAs/old mAs = new grid factor/old grid factor
Relationship between spatial resolution and relative system speed is
no relationship
Scatter carries
nonuseful information
Grid frequency
number of grid lines per mm or inch aka grid density
When the X-ray beam is not lined up with the center of the grid you have
off-center cutoff
When the SID is above or below the focal range you have
off-focus cutoff
Only grid cutoff error that can occur with a parallel grid
off-level
When the X-ray beam is angled across the lead strips you have
off-level cutoff
Material between the lead strips of a grid must be
radiolucent
Relationship between height of lead strips and distance between them is grid
ratio
During the exposure, the grid in the bucky
reciprocates, moves slightly from side to side
Field shape produced by collimator is
rectangular
Significant amount of collimation will have what effect of the amount of radiation reaching the IR?
reduced
Grid cutoff results in
reduced exposure to IR a light film-screen image possible quantum mottle
Using a grid to reduce scatter results in
shorter scale contrast and decreased IR exposure
The probability of a Compton interaction occurring is related to
the energy of the X-ray photon
Grid ratio
the height of the Pb strip to the width of the interspace
The higher the lead content,
the more efficient the scatter clean up
The numerical value assigned to each pixel is based on
the relative attenuation of x-rays transmitted through the part
Linear grids are most commonly used because
the tube can be angled in the direction of the lead lines
Quantum noise is a result of
too few x-ray photons reaching the image receptor
Grids absorb scatter because the scattered photons
travel at an angle to the transmitted photons
Adding a grid or increasing grid ratio results in an increase in patient dose
true
Grid conversion factor or bucky factor
used to determine the adjustment in mAs needed when changing grid ratios
Moire effect can be caused by
using a grid with frequency similar to CR laser scanning frequency using a grid cassette in the bucky
The amount of scatter produces is dependent on
volume of tissue exposed and kVp used
Two major factors that affect amount of scatter produced and exiting the patient are
volume of tissue irradiated and kVp
Type of grid that is usually taped to the front of an IR is
wafer grid
Moire effect is also known as
zebra pattern