Abiomed

Réussis tes devoirs et examens dès maintenant avec Quizwiz!

WILL GO TO P LEVELS IN 3 HOURS

AUTO MODE

10 x 20 mm

Abiomed recommends a _____________ hemashield platinum or vascutex graft for the impella axillary 5.0?

2-3 mls

Blue Purge Pressure Reservoir holds how many mls?

Allows the device to be turn off, but maintains purge fluid delivery, thus keep the AIC from alarming

Explain why Surgical Mode would be used?

8-12 normal > means left ventricle failure or tampnade < means hypovolemia LVEDP (VENTRICULAR PRELOAD) IS INCREASED PWP IS ALSO INCREASED

PWP

True

T/F - Only the 5.0 & LD have dual pressure sensors?

True

T/F - The 5.0 LD is the only impella with a ruler in the kit?

True

T/F- The faster the impeller rotates inside the cannula, the higher the flow produced by the Impella catheter?

26 lbs

The AIC WT is?

1 hour

The AIC battery life is?

23 fr

The Axillary 5.0 kit has a ______ peel away sheath?

Prevents blood from entering the catheter motor

The Impella System requires a purge system to:

3.5 mm

The cath inlet area is ___________below the aortic valve?

Automated Impella Controller (AIC)

The function of the _____________ is to provide an interface for monitoring and controlling the Imella cath, purge fluid to the cath, and backup power.

60-70

The impella 5.0 should be at _________ degree angle to facilitate passage of the catheter?

17 mm

The length of the LD from tip to cath is ______?

7 mm

The recommended vessel DM for the 5.0 is ______ ?

True

True or False - 5.0 LD has the impella Sponge plugs attached to the 9fr Catheter

True

True or False - The 5.0's have electrical sensor not fluid fill sensors.

True

True or False- the Impella 5.0 LD does not have a pigtail ?

1. Note placement waveform has shifted up/down on display. 2. Flow does not match the current performance level setting

Under what circumstances would you want to zero the differential pressure sensor?

2.5 = 1010

WHAT IS THE MOTOR THRESHOLD LEVEL

3.5 = 1130

WHAT IS THE MOTOR THRESHOLD LEVEL

5.0 = 1190

WHAT IS THE MOTOR THRESHOLD LEVEL

Mural Thombus in LT Ventricle Mechanical Aortic valve Aortic valve stenosis 0.6cm or less Aortic valve insufficiency graded > 2+ Access PAD

What are the Impella 2.5 contraindications?

1. < HH levels 2. Dark or blood urine 3. Acute Renal Failure

What are the s/s of hemolysis?

21 fr / 9 fr

What is the 5.0 pump and catheter size?

> 250

What is the ACT for Cath Lab?

160-180

What is the ACT range for the ICU?

2 - 30 ml/hr

What is the AIC normal purge flow rate?

300-1100 mmHG

What is the AIC normal range of the purge pressure

14 fr

What is the FR size for Impella CP

12 fr

What is the FR size for the Impella 2.5

Parasternal long axis

What is the best view position for (Transthoracic echo) TTE?

Long axis

What is the best view position for TEE?

.018

What is the guide wire for the Impella 5.0

4.0-8.0 L/min and 2.5-4.0

What is the normal CO and CI

15-25/8-15 above normal means left heart failure

What is the normal PA pressure

800-1400

What is the normal SVR

10 normal is 2-8 < hypovolemia and shock > vasodilation right side failure needs diuresis

What is the recommended CVP in ICU?

> 60

What is the recommended MAP in ICU?

93-99 cm

What is the working length of the Impella Cath

Cook 14fr / 13cm Cook 14fr / 30cm

What other sheaths can be used for the Imella 2.5?

Cook 14fr / 30cm

What other sheaths can be used for the Imella CP

BS Platinum Plus .014 BS V-18 .018

What other wires can be used for the Impella 2.5?

Reduce flow rate (P-level), Correct I/O (VOL) and improve RT heart function

What steps to take of hemolysis and low CVP,PCWP,AOP

Reduced output suction alarms elevated CVP signs of liver failure elevated pulmonary pressures (normal is 8-12 mmhg)

What the s/s of right heart failure in patients supported by the impella ?

1. Consider alternative hospital protocol. 2. Suggest contacting MEDICAL AFFAIRS for protocol

What to do if Physician decides that it is in the best interest of the patient to operate the system without heparin?

1. Whether the controller is connected to AC power. 2. Remaining battery capacity 3. Mean calculated Impella flow 4. Current purge flow 5. Calculated Impella Cath flow during systole and diastole

Which of the following types of information can you find along the BOTTOM of each screen on the AIC?

ELISA and Serotoin

HIT (heparin induce thrombocytopenia) should be verified by a positive______________ test and __________ release test?

Clinically significant hemolysis is plasma free hemoglobin greater than 40mg/dl x 2 in a 24 hour period.

Hemolysis

In the presence of adequate volume, and placement that appears in good position using a 2D echo, unresolved hemolysis requires that you continue to reposition the Impella. The rational being this is a 2D image and you cannot tell if the Impella is under Mitral apparatus.

Hemolysis

CONTRACTILITY PRELOAD- DEPENDENT AFTERLOAD- SENISTIVE

IMELLA IS PERFORMANCE IS DEPENDENT ON

First Button: MUTE ALARM Second Button ; FLOW CONTROL Third Button; DISPLAY Fourth Button ; PURGE SYSTEM Fifth Button ; MENU

Identify the buttons on the AIC ??

Dense mosaic pattern of turbulence appears above the aortic valve near the outlet area of the catheter

Describe what correct impella catheter position look like when color Doppler is added to echo?

1. Turn AIC on Hold For 3 Sec or till light alluminates 2. Press MENU and select CASE START and Press Select knob 3. Open Cassette Door by pressing the left release bottom. 4. Spike D5W bag, and insert cassette and slide purge pressure transmitter and close door 5. Plug in the red transducer plug in and snap the plastic clip to the connector cable. 7. once prime, connect red to red and yellow to yellow and make sure the conections are hand tight. 8. connect the connector cable to the connector plug on the AIC. 9. squeeze the white flush valve, and enter the purge data.

AIC Start up

Boost will return to P8 after 5 minutes

BOOST

COXMAPX.0022 IMPELLA . O.7 WATTS NORMAL IS 1-1.5

CPO CALULATION

Dense mosaic pattern beneath the aortic valve indicates that the outlet area of the catheter

Describe what incorrect impella catheter position look like when color Doppler is added to echo?

1. Low value of placement signal reduced below zero 2. Motor current normal 3. Home screen "wrong position"

Describe what you will see if the 5.0 catheter outlet area on or near aortic valve?

1. Placement signal Dampened or Flat; increased flow rate 2. Home screen "wrong position"

Describe what you will see if the 5.0 is in the ventricle or fully in aorta?

1. inadequate blood volume. 2. incorrect position within the LT Ventricle 3. RT heart failure

Describe why a suction alarm may occur?

Pulsatile placement signal Pulsatile motor current Home screen 'OK" position

Describe how the correct 5.0 AIC appears

1. Limits the amount of support 2. Decrease in arterial pressure and CO 3. Damage RBC's, leading to hemolysis

Describe the physiologic effects of prolong suction alarms?

INCREASED AFTERLOAD @ P-2/P-4 TO VENT THE LT VENTRICLE

ECMO RESULT IN


Ensembles d'études connexes

Numerical Methods 'Final' Questions

View Set

McFall AP Statistics Semester Exam

View Set

NURS 1200 Test 4 Study Guide from Sanderson - Infection (STIs, Communicable Disease, HIV)

View Set