The ideal exposure range for a digital IR is
-30% to + 50%
Diagnostic interpretation by the radiologist should be done at a ___________ display monitor. a. 1 megapixel b. 2 megapixel c. 5 megapixel
5 megapixel
If a DI system can resolve line pairs as small as 0.2 mm, what is the spatial frequency? A. 5.0 line pair/mm B. 2.0 line pair/mm C. 0.4 line pair/mm D. 0.2 line pair/mm
5.0 line pair/mm
In DI, the graphic display of pixel values and counts is the image A. Matrix B Histogram C. Bit depth
histogram
In digital imaging, higher ___ techniques can be used to reduce patient doses A. kVp B. mAs C. SID D. OID
kVp
In order for you an exposure indicator number to be accurate at least ____ of the chosen CR plate must be exposed.
30%
The DI system with a ___ pixel bit depth will have the greatest contrast resolution A. 8 B. 10 C. 12 D. 16 E. 32
32
Which of the following changes in the exposure indicator values does not represent a doubling of radiation exposure to the IR? a. S number (Fuji) changing from 200 to 400 b. S number (Fuji) changing from 400 to 200 c. EI (Kodak/Carestream) changing from 1200 to 1500 d. EI (Kodak/Carestream) changing from 1500 to 1800
A
Define auto-rescaling
Adjustment of the electric signal to compensate for exposure error as determined by the histogram analysis.
Mathematical calculations used for DI construction are called A. logarithms B. algorithms C. interpolations D. extrapolations
Algorithms
The distracting area of increased brightness surrounding a collimated image (primary beam restriction) can be removed from the displayed digital image by: a. Black masking b. Shuttering c. Electronic collimation d. Black border e. All of the above
All of the above
The histogram generated from image data allows the system to find the A. Minimum signal (S1) within the VOI B. Maximum signal (S2) within the VOI C. Relative distribution of pixel values D. All of the above
All the above
In CR, the image must be converted into digital form by what device? A. Digital to analog converter B. Flux capacitor C. Analog to digital convertor D. DVD-ROM
Analog to digital convertor
Soft copy viewing means that the image is being viewed: a. At a computer b. On a soft piece of film c. In a soft light environment d. None of the above
At a comupter
Based on the histogram analysis, consistent image brightness is accomplished through: a. Adjustment of window width b. Automatic rescaling c. Manipulation of the exposure indicator d. Manipulation of the LUT
Automatic rescaling
CR plate phosphor crystals are most often___ with europium activators 1. BaSO4 2. BaFBr 3.BaFI 4. AgI
BaFBr & BaFl
CR plates are erased every morning mainly to eliminate the effect of A. Background radiation B. Scatter radiation C. Light leaks into the cassette D. Computer noise
Background radiation
Window level in digital imaging controls what image display property? A. Brightness B. Contrast C. Resolution D. Distortion
Brightness
A photostimulable phosphor is the active component in what piece of equipment? A. Intensifying screen B. Computed radiography imagining plate C. AEC chamber D. Fluroscopic screen
Computed radiography imagining plate
Window width in digital imaging controls what image display property? A. Brightness B. Contrast C. Resolution D. Distortion
Contrast
Boxes of laser film should be stored in a A. cool, humid area B. cool, dry area C. warm, humid area D. warm, dry area
Cool, dry area
When utilizing a DR system, if 2x the appropriate mAs is used, the result will be an image with ____ brightness and ___ contrast A. Correct, correct B. Excessive, correct C. Correct, low D. Insufficient, low
Correct, Correct
What is shuttering
Cropping or A workstation software function allows a black mask to be placed over bright areas.
The communication standard for medical images is: a. PACS b. HL7 c. DICOM d. RIS
DICOM
An IR that has a higher ___ has a higher probability of interacting with exit radiation (A)LET (B)TFT (C)SNR (D)DQE
DQE
What is used to measure an IR system's x-ray absorption efficiency? (A) Fill factor (B) Exposure data recognition (C) Detective quantum efficiency (D) Conversion efficiency
DQE ( detective quantum efficiency)
Increasing the window width of a DI while viewing it on the monitor will A. Increase image brightness B. Decrease image brightness C. Increases image contrast D. Decrease image contrast
Decrease image contrast
__________ is defined as the measure of x-ray absorption efficiency of an IR.
Detective quantum efficiency (DQE)
What type of DR system utilizes a-Seas a component of the detector element? (A) Direct conversion (B) Indirect conversion
Direct conversion
The readout obtained after an exposure is taken with DR equipment that indicates patient exposure is the _____
Dose are product (DPA)
The range of exposures over which an IR can adequately acquire image data is called the ____ range. A. Focal B. Pixel depth C. Dynamic D. Acquisition
Dynamic
This postprocessing technique improves the visibility of small, high contrast structures. a. Smoothing b. Contrast enhancement c. Edge enhancement d. Subtraction
Edge enhancement
What DI post-processing technique improves the visibility of small, high contrast structures? (A) Edge enhancement (B) Smoothing (C) Contrast enhancement (D) Subtraction
Edge enhancement
What postprocessing technique would be helpful in radiography of the AP thoracic spine?
Equalization
The exposure to a digital IR is represented by the numerical value called a/an A. Algorithm B. Exposure index C. Histogram D. Signal to noise ratio
Exposure index
The numerical value indicating the level of radiation exposure to the digital IR is the: a. VOI b. Histogram c. Exposure indicator d. LUT
Exposure indicator
In an indirect DR capture system, the absorbed x-ray photons are first converted to A. Fluorescent light B. Electrons C. Heat
Fluorescent light
What type of laser is used to scan the CR IPs? A. Fluorohalide B. Argon C. Helium-neon D. Cesium
Helium-neon
A graphic display that is constructed in digital imaging to show the radiographer the distribution of pixel values is a A. Algorithm B. Histogram C. Voxel D. Matrix
Histogram
The graphic display of pixel values is the _____
Histogram
How are pixel size and spatial resolution related? A. Directly- larger pixels = greater resolution B. Inversely- smaller pixels = greater resolution
INVERSLEY: Smaller the pixels=greater resolution
This post-processing function reverses the original gray scale values, changing the image from a negative to a positive one. a. equalization. b. flat-fielding. c. image reversal. d. subtraction.
Image reversal
Which of the following would reduce quantum noise in a digital image? 1. Decreasing mAs 2. Increasing mAs 3. Changing to a lower ratio grid a. 1 & 2 only b. 1 & 3 only c. 2 & 3 only d. 1, 2, & 3
Increasing mAs & Changing to a lower ratio grid
The type of DR detector that utilizes CSI and amorphous silicon is a ___ conversion A. Direct B. Indirect
Indirect
What type of IR system has the greatest DQE?
Indirect a:silicon flat panel DR detectors
CR plate erasure is accomplished by A. Light B. Electrons C. Heat D. Thermionic emission
Light
LCD stands for: a. Liquid cathode display b. Liquid crystal display c. Logarithmically created digitization d. Layered components diagram
Liquid crystal display
Describe the effect of kVp level on the width of the histogram.
Low kV gives a wider histogram, high kV gives a narrower histogram.
A digital image is composed of rows and columns known as a A. Matrix B. Pixel C. Voxel D. Histogram
Matrix
Compared with F/S IRs, digital receptors absorb ___ scatter radiation A. More B. Less C. The same quantity
More
The computer system designed to distribute, store, and display digital images is the _____
PACS
In the AUTO processing mode of a CR system how must the part and collimation edges be related to the IP?
Part and beam must be centered to the IP and even borders of collimation must be present.
______ occurs during CR plate processing when the trapped electrons are returned to a lower energy state. A. Luminescence B. Thermionic emission C. Analog to digital conversion D. Photostimulated luminescence
Photostimulated luminescence
A ______ is the smallest component of a digital image (A)Pixel (B)Voxel (C)Matrix
Pixel
___ is defined as the distance between the center of a pixel to the center of the adjacent pixel in a DR system DEL. A. Bit depth B. Window level C. Window width D. Pixel pitch
Pixel pitch
Adjustment of the digital signal to compensate for exposure error is called (A)Smoothing (B)Rescaling (C)Fixation (D)Enhancement
Rescaling
The minimum useful signal on a histogram is termed __
S1
The maximum useful signal on a histogram is termed ___
S2
In DR systems what parameters determine which DELs will be included in the image processing.
Selected FOV and collimation
How does underexposure affect the SNR and image quality?
Since the signal is very low, there is excessive noise compared to the useful signal. Image quality is low and the image demonstrates quantum mottle.
The postprocessing technique that is used to suppress image noise is called A.Smoothing B. Edge enhancement C. Contrast enhancement D. Subtraction
Smoothing
This postprocessing technique reduces the visibility of image noise. a. Smoothing b. Contrast enhancement c. Edge enhancement d. Subtraction
Smoothing
_ resolution is the ability of the system to display small high contrast structures accurately. A. Spatial B. Contrast
Spatial
This postprocessing technique can eliminate structures superimposing the structures of interest.
Subtraction
What DI post-processing function removes superimposing structures from the image? (A) Edge enhancement (B) Contrast enhancement (C) Black/white reversal (D) Subtraction
Subtraction
The relationship between useful and useless information an IR receives is called A. The fill factor B The DQE C. The SNR D. The algorithm
The SNR
Define Modulation Transfer Function (MTF)
The ability of a system to record spatial frequencies
The % of a flat panel detector's DEL that is able to absorb radiation (is the actual detector) is called (A ) the fill factor (B) the DQE (C) the SNR (D) the algorithm
The fill factor
How is the display matrix size related to image resolution?
The greater the matrix size, the greater the resolution
What compromise is made by a department that chooses to utilize only hard copy viewing of images?
The inability to utilize the dynamic range, cannot window and level.
Menu selection of the anatomical part and view chooses the algorithm- define algorithm.
The mathematical formula used to construct the histogram
The CR reader component that converts the released light into electrons is the (A)Beam shaping optics (B)The ADC (C)The laser beam (D)The photomultiplier tube
The photomultiplier tube
Define Exposure Field Recognition.
The process in CR where the system first scans the entire plate in order to locate the clinically useful information.
Define signal to noise ratio (SNR).
The relationship between useful information (signal) representing anatomy and useless information (noise).
What is represented on the vertical (y) axis of the histogram?
The total number of each specific pixel value
What is represented on the horizontal (x) axis of the histogram?
The various pixel values (shade of gray
Most laser film must be handled A. under a Wratten 6B safelight B. in total darkness C. under a GBX safelight D. with high-temperature chemical processors
Total darkness
Direct digital radiography systems eliminate the EFR step. True / False
True
Dry laser printers use heat to print the image. A. True B. False
True
DI image noise results mostly from A. Overexposure B. Underexposure C. Use of low ratio grids D. Excessive OID
Underexposure
In digital imaging, quantum mottle is a result of A. underexposure B. overexposure C. histogram error D. incorrect collimation
Underexposure
What choice can you make as an imager in cassette selection when utilizing a CR system to produce the greatest spatial resolution?
Use the smallest size CR plate that will include all the structures of interest.
During laser scanning of a CR IP, the stored energy is released in the form of A. Electrons B. Visible blue light C. visible red light D. Heat
Visible blue light
What is image stitching
When anatomy or the area of interest is too large to fit on 1 cassette, multiple images can be "stitched" together using specialized software programs
A change in the __________ will result in an image with a change in image brightness.
Window level
This post-processing function allows the viewer to adjust the image brightness and/or contrast. a. edge enhancement b. image reversal c. subtraction d. windowing
Windowing
The SI unit used to specify DAP is the
dGy/cm2
The AAPM has proposed adoption of a universal exposure indicator value called the_____ _____.
deviation index
Which of the following is not accomplished through the histogram analysis? a. Compensation for image over or underexposure b. Identification of borders of collimation c. Comparison with a LUT d. Determination of S1 and S2 values e. All of the above occur during histogram analysis
e
Higher DQE image receptors can result in ________ patient doses. A. Higher B. Lower
lower
Increasing the window width will result in an image with: a. Lower contrast b. Higher contrast c. Decreased brightness d. Increased brightness
lower contrast
The unit of spatial frequency that is used to specify spatial resolution is A. micrometers B. MTF C. lp/mm D. Hertz
lp/mm