RAD 274 Quiz 3
Some Advantages of digital imaging over conventional screen/film radiography include: 1. immediate visualization of the image 2. ability to adjust the image post exposure 3. highest spatial resolution possible
1,2,& 3
Which of the following is a contributor to image noise? 1. Scatter radiation 2. plus density artifacts 3. minus densities artifacts
1,2,&3
What is the pixel size in millimeters of a 256 matrix with a 25 cm field of view?
1.0 mm Pixel size= FOV / matrix
Of the following: Which matrix size has the smallest pixels? 128 x 128 256 x 256 512 x 512 1024 x 1024
1024 x 1024
The device that is needed to allow computer processing of an analog signal is a/an:
ADC (analog to digital converter)
Contrast resolution in a digital image: A. Is the ability of a imaging system to display a short scale of contrast B. Is the ability of a system to display a large number of shades of gray C. Is the ability of an imaging system to display very bright pixels D. Is controlled by the number of pixels and pixel bit depth
BOTH B&C
Contrast resolution in a digital image: A. Is the ability of a imaging system to always display a short scale of contrast B. Is the ability of a system to display a large number of shades of gray C. Is the ability of an imaging system to display very bright pixels D. Is controlled by the number of pixels and pixel bit depth
BOTH B&D
if FOV decreases and the Matrix size remains the same: Pixel size will _____ and spatial resolution will _____
Decrease, Increase
if FOV remains constant and the matrix size increases: Pixel size will _____ and spatial resolution will _____
Decrease, increase
if FOV increases and Matrix remains the same: Pixel size will _____ and spatial resolution will _____
Increase, Decrease
If FOV remains constant and the matrix size Decreases: Pixel size will _____ and spatial resolution will _____
Increase, decrease
The latent image in a CR imaging plate is stored by the:
The latent image in a CR imaging plate is stored by the:
In order to view a digital image that has been exposed on a phosphor plate cassette, what "processing" equipment is needed?
a laser reader
A digital matrix is made up of:
a matrix of pixels
The IR Exposure response curve for digital imaging is linear rather than S-shaped because:
a wider dynamic range
Any unwanted information on an image is the definition for:
artifacts
The active layer in the photostimulable phosphor of a CR imaging plate is typically made of:
barium fluorohalide
language used by computers
binary
calculated using the formula 2n; reports the number of gray shades available
bit debth
The number of shades of gray that can be demonstrated in a digital image is affected by:
bit depth
The amount of luminance of an image or light displayed by a monitor is termed:
brightness
The amount of luminance of an image or link displayed by a monitor is termed:
brightness
The window level control is used to alter image:
brightness
amount of luminance an image displays on a monitor
brightness
If a pixel has a bit depth of 8, this means that the pixel: ( hint--- 2n)
can produce 256 shades of gray
The window width control is used to alter image:
contrast
resolution; the ability of a system to distinguish between and/or display objects that attenuate x-ray photons differently
contrast
In general, as image noise increases:
contrast decreases
If pixel pitch is increased, the image will have/display:
decreased spatial resolution
the ability to detect and display a large range of radiographic densities: a wide ______
dynamic range
Before a computed radiography imaging plate can be reused, it must be:
erased with a high intensity light
portion of the collimated field that contains the anatomy and displayed on the monitor
field of view
On a PA projection of the wrist, you notice what appears to be a soft tissue mass superimposed over the bony anatomy in the digital DR image. When you inspect the patient, you notice the presence of a large linear scar from a previous injury. The visualization of this scar in the image is primarily due to what characteristic of digital detectors?
high contrast resolution
The field of view (FOV) best describes:
how much of the patient is imaged in a matrix
One of the advantages of digital imaging is: -several visual cues are present that indicate over or under exposure of the image - images can be easily sent to remote locations - the image is permanently fixed, preventing post-exposure adjustments - it represents a continuous and analog range of signal
images can be easily send to remote locations
If the field of view remains unchanged AND the matrix size increases, then the resulting spatial resolution will be _____ because the pixel size will decrease.
improved
IF the field of view (FOV) increases AND the matrix size is unchanged, then the resulting pixel size MUST ________ and as a result the spatial resolution will decrease.
increase
Which of the following image quality changes is undesirable? - increased spatial resolution - increased contrast resolution - increased image noise - increased image matrix size
increased image noise
If pixel density is increased, the image will have/display: - increased spatial resolution - decreased spatial resolution - increased contrast resolution - decreased contrast resolution
increased spatial resolution
Spatial detail (resolution) in a digital imaging system would be improved MOST by: - increasing window width - decreasing window level - increasing the number of pixels - decreasing the flux gain of the photomultiplier tube
increasing the number of pixels
Which of the following will allow more line pairs per millimeter to be imaged? - using a smaller matrix - increasing the number of pixels in the matrix - increasing the bit depth - using a higher frequency signal
increasing the number of pixels in the matrix
The purpose of europium "doped" in the photostimulable phosphor is to:
it acts to trap(store) the electrons (metastable e-)
array of cells arranged in columns and rows
matrix
If the window width control is opened wider (a larger range):
more gray shades will be available
If the window width control is opened wider:
more gray shades will be available
image quality constituted of unwanted or useless background information
noise
The device that converts the light emitted by an exposed PSP inside the computed radiography (CR) reader into an electric signal is the:
photodiode (photodetector)
each cell in a matrix is called
pixel
the distance of spacing from the middle of one pixel to the middle of an adjacent pixel
pixel pitch
The use of high-resolution monitors: (High resolution means a monitor with a large matrix size)
promotes or improves the visibility of anatomical structures
The process of assigning number values to each pixel in proportion to the energy received best describes:
quantization
assigns a value to each pixel relative to the energy it received
quantization
The anatomical structures that would cause the greatest degree of x-ray photon attenuation and image brightness would be those that are:
radiopaque
If a high degree of quantum mottle is noted on an image, the most appropriate solution to correct this exposure error would be to:
repeat the exposure using a higher mAs
process of detecting and measuring exposure received by the detector
sampling
process of dividing image up into a matrix, each cell becomes a pixel
scanning
The laser in a CR reader...
scans the exposed phosphor in a raster pattern
formula to calculate this is 1/2 (PS); determines the number of line pairs that can be resolved if the pixel size is known
spatial frequency
After exposure, the remaining steps for acquiring a computed radiography image, in order, are:
stimulation, reading, and erasing
the minimum number of pixels needed to image one line pair
two
postprocessing function that alters image brightness
window level
postprocessing function that alters image contrast or grayscale
window width