SPECT
A source that is used for uniformity correction on a SPECT camera should have no more that how much variation in uniformity?
(+/-) 1%
How many counts should be acquired for a uniformity correction flood on a 128 x 128 matrix?
120 million
How many frames are required to evaluate an ejection fraction on a MUGA study?
16
SPECT data may be acquired in a continuous-motion orbit or a step-and-shoot method. The step-and-shoot method adds how much time per projection?
3-5 seconds
If arrhythmias reach __________ percent, it is generally advisable not to analyze the data.
30
If you go from a 64 x 64 matrix to a 128 x 128 matrix size, by what factor are you increasing the number of pixels?
4
A SPECT is set up to do a 360 degree rotation around a patient.The parameters are for 64 stops with 30 seconds per stop.What is the most likely matrix size that should be used?
64x64
If an image is acquired into a 64 x 64 matrix on a scintillation camera with a 496.26 mm diameter field of view, what are the dimensions of each pixel?
7.75
How many frames are acquired during a typical List Mode acquisition?
800-1200
Which reconstruction method typically demonstrates poor, artifact ridden images?
Backprojection
The primary benefit of SPECT is:
Improved contrast
What could you do as a technologist to help eliminate the star artifact seen on reconstructed SPECT images?
Increase the number of stops
Which of the following is not appropriate when setting up a MUGA acquisition?
Multi-crystal camera
What is the term used for the count profile of the rays that compose an image?
Projection
Pixel width determines which of the following?
Resolution
What is the reason that SPECT imaging significantly improves the image quality over planar imaging?
SPECT removes out of plane information
Which type of filter will eliminate noise from a gated study?
Spatial filter
The Jaszczak phantom tests for ___________________________ on a SPECT camera.
Spatial resolution
The ability of a system to distinguish two objects as separate objects is known as what?
Spatial resolution
The order of the butterworth filter controls what?
The point at which the slope hits the baseline
What is the name of the artifact you will receive when the center or rotation correction has not been done for a 180 degree rotation?
Tuning fork artifact
Iterative reconstruction algorithms require more time to process than FBP but produce
images with less noise